[Thyroid uptake error in four different gamma camera systems].

Abstract

The thyroid uptake of (123)I and (131)I is generally measured by a gamma camera system. We evaluated the error in determining thyroid uptake caused by different methods of calculation among four gamma camera systems with various collimators. We first designed an original thyroid phantom that consisted of the thyroid and a body containing various levels of radioiodine activity. The applications for thyroid uptake equipped in two gamma camera systems performed calculations by the automatic method with background counts not subtracted from the capsule counts. When the size of the rectangular region of interest (ROI) for the capsule was set at 10x8 cm (a typical ROI size for the thyroid), percentages of thyroid uptake as calculated by the manual method with background counts subtracted from the capsule counts and thyroid counts were 52% to 57% when the value was set at 55% for (123)I; and 54.2% and 58.7%, respectively, when the value was set at 60% for (131)I. On the other hand, the percentages of thyroid uptake calculated by the automatic method with the application using two gamma camera systems with non-subtraction of background counts from the capsule counts were 46% and 50.5% when the value was set at 55%; and 49.6% when the value was set at 60%. The values calculated by the automatic method were underestimated as a result of background counts that were not subtracted from the capsule counts. When ROI size for the capsule was set at 4x4 cm, which is slightly larger than the capsule size, even thyroid uptake as determined by the automatic method using the application showed a difference of less 2% from the set values. There was no difference in thyroid uptake among the various kinds of collimators, high-resolution collimators, all-purpose collimators, and a suitable collimator for gamma-ray energy of (123)I.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Ai Yoshida, Kyoko Saito, Yoshimune Endo, Kazuaki Mori, Misayo Seki, Hidenori Takebe. 2004. [Thyroid uptake error in four different gamma camera systems].. https://doi.org/10.6009/jjrt.kj00000922342

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Radioguided sentinel lymph node biopsy in patients with malignant cutaneous melanoma: the nuclear medicine contribution.

As for other solid tumors, malignant cutaneous melanoma drains in a logical way through the lymphatic system, from the first to subsequent levels. Therefore, the first lymph node encountered (the sentinel node) will most likely be the first to be affected by metastasis, and a negative sentinel node makes it highly unlikely that other nodes in the same lymphatic basin are affected. Sentinel lymph node biopsy distinguishes patients without nodal metastases, who can avoid nodal basin dissection with its associated risk of lymphedema, and those with metastatic involvement who might benefit from additional therapy. This procedure represents a significant advantage as a minimally invasive procedure, considering that only an average 20% of melanoma patients with Breslow thickness between 1.5 and 4 mm harbour metastasis in their sentinel node(s) and are therefore candidates to elective lymph node dissection procedures. The cells that originate cutaneous melanomas are located between dermis and epidermis, a zone that drains to the inner lymphatic network in the reticular dermis, in turn to larger collecting lymphatics in subcutis. Therefore, the optimal modality of interstitial administration of radiocolloids for lymphoscintigraphy and subsequent radioguided sentinel lymph node biopsy is through intradermal/subdermal injection. (99m)Tc-labeled colloids in various size ranges are equally adequate for radioguided sentinel lymph node biopsy in patients with cutaneous melanoma, depending on local experience and availability. For melanomas located in the midline area of the head, neck, and trunk, particular consideration should be given to ambiguous lymphatic drainage, which frequently requires interstitial administration virtually all around the tumor or surgical scar from prior excision of the melanoma. Lymphoscintigraphy is an essential part of radioguided sentinel lymph node biopsy because images are used to direct the surgeon to the sites of the node(s). The sentinel lymph node should have a significantly higher count than that of background (at least 10:1 intraoperatively). After removal of the sentinel node, the surgical bed must be reexamined to ensure that all radioactive sites are identified and removed for analysis. The success rate of radioguidance in localizing the sentinel lymph node in melanoma patients is about 98% in institutions where a high number of procedures are performed, approaching 99% when combined with the vital blue dye technique. The procedure is becoming the standard of care for patients with cutaneous melanoma because of its high prognostic value that has led to include the procedure in the most recent version of the TNM staging system.

Gamma Cameras↗

[Dependence of uniformity on the radionuclide in SPECT: test methods].

The aim of this study was to investigate test methods to clarify whether the non-uniformity of a gamma camera depends on individual radionuclides, and whether it is necessary to measure a separate correction matrix for each radionuclide used in single photon emission computed tomography (SPECT). Two methods were devised to verify the nuclide-dependence of the gamma camera. In order to test the energy correction of the detectors, the first approach was based on the evaluation of the intrinsic non-uniformity and on the production of images with asymmetrical energy window. The second method was based on the production of correction matrices for different radionuclides, as well as on the subsequent application to phantom data that were also generated with different radionuclides. The investigation of a dualhead gamma camera produced the same results with both methods. One detector head was found to be weakly dependent on the radionuclide, due to the insufficient quality of energy correction. In this case, the phantom or patient data should be corrected using a uniformity correction matrix measured with the same radionuclide. The second detector remained nuclide-independent; in this case the uniformity correction matrix acquired for only one radionuclide was sufficient.

Gamma Cameras↗

New devices for imaging in nuclear medicine.

Pinhole gamma camera imaging offers the ability to obtain high resolution images from single gamma ray emitting radiotracers playing a reasonable tradeoff between very small field of view (FoV) and sensitivity. On the other hand the total spatial resolution of a pinhole imaging device is predominantly affected by the detector intrinsic spatial resolution for reduced magnification factors. To design very compact pinhole SPET scanners with very high intrinsic spatial resolution, authors investigated a miniature gamma camera based on the newly developed Hamamatsu H8500 flat panel photomultiplier. The PSPMT was coupled to the following scintillation arrays: CsI(Tl) array with 0.2-mm, 1-mm, 1.4-mm pixel size and NaI (Tl) with 1-mm pixel size. The imaging performances were evaluated by 57Co spot and flood irradiations. NaI(Tl) array shows a better pixel identification for 1 mm pixel size, proving to be a good candidate to make a large area photodetector based on multi PSPMTs closely packed. Although CsI(Tl) array had the smallest pixel size, the low light output limited the best intrinsic spatial resolution to about 0.5 mm.

Gamma Cameras↗