PubMed Health⌕ Search

PubMed · 15320462

Reducing power line interference in digitised electromyogram recordings by spectrum interpolation.

Abstract

Interference from power lines (50 or 60 Hz) is the largest source of extraneous noise in many bio-electric signals and is within the bandwidth of many such signals. In this study, two different methods were compared for their efficacy in removing 50 Hz noise added to surface electromyogram (EMG) signals free of power line interference. The first was a simple second-order recursive digital notch filter. The second was an approach called spectrum interpolation, in which it is assumed that the magnitude of the original 50 Hz component of the EMG signal can be approximated by interpolation of the amplitude spectrum of the signal. When the spectrum was based on records containing an integer number of cycles of 50 Hz interference, and the frequency resolution was finer than 1 Hz, spectrum interpolation performed similarly to, or significantly better than, the notch filter (p < 0.01). It was also possible to make spectrum interpolation more robust than the notch filter. The Pearson squared correlation coefficient r2 between clean signals and signals processed using the notch filter was reduced from 0.98 to 0.65 when the interference frequency was increased by 0.5 Hz, but r2 for spectrum interpolation at 0.2 Hz resolution was only reduced from 0.99 to 0.85 if spectral values between approximately 49.5 and 50.5 Hz were modified by interpolation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D T Mewett, K J Reynolds, H Nazeran. 2004. Reducing power line interference in digitised electromyogram recordings by spectrum interpolation.. https://doi.org/10.1007/bf02350994

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Striping artifact removal in VisiumHD data through nuclear counts modeling.

MOTIVATION: 10x Genomics VisiumHD enables spatial transcriptomics at 2&#x2009;&#xb5;m &#xd7; 2&#x2009;&#xb5;m resolution but exhibits slide-specific, non-periodic striping artifacts due to lane-width variability. These multiplicative row/column effects distort bin total counts and can bias downstream analyses. The state-of-the-art destriping approach is the normalization procedure used as a preprocessing step in bin2cell; it applies sequential high-quantile row- then column-wise normalization, which is asymmetric and can introduce edge effects/macro-stripes and distortions of large-scale total-count structure. RESULTS: We propose a statistical destriping approach that leverages nuclei segmentation from the co-registered H&E image. Assuming transcript abundance is constant within each nucleus, we model bin counts with a negative binomial distribution whose mean is a product of a nucleus-specific concentration and row- and column-specific stripe-factors reflecting lane-width variation. We fit all parameters in a generalized linear modeling framework with cross-validated regularization on stripe-factors and iterative dispersion estimation, and use the fitted parameters to correct the observed counts into a destriped image. On synthetic data with known ground truth, our method improves stripe-factor estimation accuracy and reduces error in corrected counts relative to bin2cell and bin2cell-derived baselines. Across four public VisiumHD slides, it consistently lowers striping intensity while substantially better preserving biological signal present in the large-scale global count structure and avoiding the artifacts introduced by other methods. AVAILABILITY AND IMPLEMENTATION: All source code and links to publicly available data used for this study are available at https://github.com/paolamalsot/destriping-GLM.

Artifacts↗

Impairment of echocardiographic acoustic window caused by breast implants.

Cosmetic breast implants are increasing in popularity. The presence of a breast implant overlying the anterior mediastinal space as a cause of impairment of the echocardiographic acoustic window has not been described previously. Here, we report three cases with significant impairment of echocardiographic acoustic window caused by breast implants. Clinicians should be aware of this interference and women should be informed of this dilemma before considering this cosmetic surgery.

Artifacts↗