PubMed Health⌕ Search

PubMed · 15626601

Microchamber array based DNA quantification and specific sequence detection from a single copy via PCR in nanoliter volumes.

Abstract

A novel method for DNA quantification and specific sequence detection in a highly integrated silicon microchamber array is described. Polymerase chain reaction (PCR) mixture of only 40 nL volume could be introduced precisely into each chamber of the mineral oil layer coated microarray by using a nanoliter dispensing system. The elimination of carry-over and cross-contamination between microchambers, and multiple DNA amplification and detection by TaqMan chemistry were demonstrated, for the first time, by using our system. Five different gene targets, related to Escherichia coli were amplified and detected simultaneously on the same chip by using DNA from three different serotypes as the templates. The conventional method of DNA quantification, which depends on the real-time monitoring of variations in fluorescence intensity, was not applied to our system, instead a simple method was established. Counting the number of the microchambers with a high fluorescence signal as a consequence of TaqMan PCR provided the precise quantification of trace amounts of DNA. The initial DNA concentration for Rhesus D (RhD) gene in each microchamber was ranged from 0.4 to 12 copies, and quantification was achieved by observing the changes in the released fluorescence signals of the microchambers on the chip. DNA target could be detected as small as 0.4 copies. The amplified DNA was detected with a CCD camera built-in to a fluorescence microscope, and also evaluated by a DNA microarray scanner with associated software. This simple method of counting the high fluorescence signal released in microchambers as a consequence of TaqMan PCR was further integrated with a portable miniaturized thermal cycler unit. Such a small device is surely a strong candidate for low-cost DNA amplification, and detected as little as 0.4 copies of target DNA.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Yasutaka Matsubara, Kagan Kerman, Masaaki Kobayashi, Shouhei Yamamura, Yasutaka Morita, Eiichi Tamiya. 2005-02-15. Microchamber array based DNA quantification and specific sequence detection from a single copy via PCR in nanoliter volumes.. https://doi.org/10.1016/j.bios.2004.07.002

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

From host response to genomic targets: electrochemical biosensing of tuberculosis biomarkers.

Tuberculosis (TB) remains one of the leading causes of death from a single infectious agent worldwide, with timely diagnosis continuing to be a major challenge, particularly in resource-limited settings. Conventional TB diagnostic methods are limited by low sensitivity, long turnaround times, and an inability to reliably differentiate latent from active disease. Biomarker-based diagnostic strategies have therefore gained increasing attention as they offer the potential to improve early detection, disease differentiation, and treatment monitoring. Herein, we examine electrochemical biosensing strategies for TB diagnostics using a biomarker-class-driven framework, covering host-response biomarkers (IFN-γ and TNF-α), pathogen-derived antigens (ESAT6, CFP10, CFP10-ESAT6, MPT64, Ag85, HspX and LpqH), cell-wall signatures and whole-cell markers (LAM and whole cell Mtb), and genomic markers (Mtb DNA and IS6110). Through structured comparison of recognition elements, biointerface designs, signal amplification strategies, electrochemical techniques, matrices, and validation levels, this review identifies the most promising technical approaches for different TB biomarker classes. It further highlights key translational bottlenecks, including limited clinical validation, buffer-based testing, complex multistep amplification, redox-probe dependence, matrix fouling, and insufficient evidence of manufacturability. This review therefore provides practical guidance for developing electrochemical TB biosensors that are analytically sensitive, clinically relevant, and suitable for decentralized diagnostic applications.

Biosensing Techniques↗

Metal-organic frameworks nanozyme-integrated portable microneedle patch for visual bacterial monitoring in meat.

Foodborne microbial contamination is a major global health concern, with conventional methods often being time-consuming and complex. Herein, we developed a novel portable biosensor by integrating microneedle patch technology and a metal-organic framework (Fe/Cu-NBDC MOF) nanozyme, enabling rapid, on-site, visual detection of bacteria in meat. The sensing system works by encapsulating aptamer-functionalized MOF nanozymes within a hydrogel patch, where their catalytic sites are initially blocked by the aptamer. In the presence of Staphylococcus aureus (S. aureus) as the target, the specific aptamer's binding to bacteria exposes numerous catalytic sites, further activating the chromogenic reaction of the tetramethylbenzidine‑hydrogen peroxide (TMB-H₂O₂) system, enabling visual detection of S. aureus. The biosensor demonstrates a detection limit of 82 CFU/mL with excellent specificity to successfully apply to commercial mutton. By integrating sampling, enrichment, and visual detection into a single compact device, this platform offers a practical, efficient solution for rapid on-site screening of foodborne pathogens.

Biosensing Techniques↗

The biosensor based on the pyruvate oxidase modified conducting polymer for phosphate ions determinations.

An enzymatic biosensor was fabricated by the covalent immobilization of pyruvate oxidase (PyO) onto the nano-particle comprised poly-5,2':5',2''-terthiophene-3'-carboxylic acid, poly-TTCA (nano-CP) layers on a glassy carbon electrode (GCE) for the amperometric detection of the phosphate ions. The direct electron transfer reaction of the immobilized PyO onto the nano-CP layers was investigated and the electron transfer rate constant was determined to be 0.65 s(-1). The electrochemically prepared nano-CP lowered the oxidation potential (+0.40 V versus Ag/AgCl) of an enzymatically generated H(2)O(2) by PyO in a phosphate solution. Experimental parameters affecting the sensitivity of the biosensors, such as amounts of the cofactors, the pH, the applied potential, and the temperature were optimized. A linear response for the detection of the phosphate ion was observed between 1.0 microM and 100 microM and the detection limit was determined to be about 0.3 microM. The response time of the biosensors was about 6s. The biosensor showed good selectivity towards other interfering anions. The long-term storage stability of the phosphate biosensor was studied and the sensor was applied in a human serum sample for the phosphate ions detection.

Biosensing Techniques↗