Fixation methods for implantable port chamber: comparative study using glue, self-stabilizing leg and suture fixations in rabbits.

Abstract

OBJECTIVE: To evaluate the fixation strength and tissue reaction of the glue fixation and self-stabilizing leg fixation methods and to compare the results with those of the conventional tagging suture fixation method. MATERIALS AND METHODS: Twelve healthy rabbits were selected and three different methods of implanting the port chamber were employed on the back of each rabbit. A total of thirty six port chambers were implanted with these three different methods, viz. the glue fixation method using tissue adhesive, the self-stabilizing leg method using a self-expandable stabilizing leg, and the suture fixation method. The fixation strength and the gross and histopathologic changes of each fixation method were evaluated at three days, one week, two weeks and four weeks after port implantation. RESULTS: The glue fixation method showed a good fixation strength, which was similar to that of the tagging suture method (p = 0.3486). Five of the six ports (83%) implanted with the glue fixation method which were examined after two weeks showed cracks on the external surface, but this had no adverse effects on their function. A large amount of granulation tissue reaction was found at the bottom of the chamber (p = 0.0025). The fixation with the self-stabilizing leg showed relatively lower fixation strength (p = 0.0043), but no turning-over of the chamber occurred. The fixation strength improved with time after the first week, and minimal granulation tissue reaction was observed with this method. CONCLUSION: The glue fixation method exhibited equal fixation strength compared to the suture fixation, but showed cracking and a large amount of granulation tissue, whereas the fixation with a self-stabilizing leg showed weaker fixation strength.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Hyoung Il Na, Hyung Jin Shim, Byung Kook Kwak, Hyeon Joo Kim, Yong Cheol Lee. Fixation methods for implantable port chamber: comparative study using glue, self-stabilizing leg and suture fixations in rabbits.. https://doi.org/10.3348/kjr.2004.5.4.266

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Electrochemical characterization of albumin protein on Ti-6AL-4V alloy immersed in a simulated plasma solution.

The effect of oxygen and albumin on the electrochemical behavior of a Ti-6Al-4V alloy immersed in a simulated inorganic plasma (SIP) solution was studied with a rotating-cylindrical electrode configuration to focus on the surface/electrolyte reactions. Potentiokinetic scans and electrochemical impedance spectroscopy have been used to characterize the interface by determining the passive current density and capacitance. For the polarization scans, an albumin addition of 37.7 mg/cm(3) to the SIP solution (oxygenated and unoxygenated) decreased the passive current density, indicating a lowering of the corrosive rate. The surface capacitance for the Ti-6Al-4V alloy immersed in a SIP solution averaged 13 microF/cm(2), which transformed after albumin addition (37.7 mg/cm(3)) from a potential independent behavior to the capacitance ranging from 23 to 6 microF/cm(2) with increasing potentials from -800 to 1500 mV(SCE), respectively, indicative of albumin adsorption. Within the same potential range and albumin addition to oxygenated solutions, the capacitances expanded slightly with a similar decreasing trend from 31 to 6 microF/cm(2), although the capacitance depicts an interaction between the hydrated passive film and the adsorbed albumin from -550 to 500 mV(SCE) in which the capacitance plateaued at 15 microF/cm(2). The hydrated porous oxide film results from the porous rutile layer reacting with H(2)O(2) formed as an intermediary component of oxygen reduction at the Ti-6Al-4V surface. The passive film-albumin interaction would affect the processing of titanium alloys in their surface preparation for biocompatibility, as well as determining the reactivity of titanium alloys to proteins.

Alloys↗

Fatigue-crack growth properties of thin-walled superelastic austenitic Nitinol tube for endovascular stents.

Over the past 10 years, the supereleastic nickel-titanium alloy Nitinol has found widespread application in the manufacture of small-scale biomedical devices, such as self-expanding endovascular stents. Although conventional stress/strain-life (S/N) analyses are invariably used as the primary method for design against fatigue loading and for predicting safe lifetimes, fracture mechanics-based methodologies provide a vital means of assessing the quantitative effect of defects on such lifetimes. Unfortunately, fracture mechanics studies on fatigue in Nitinol are scarce, and most results do not pertain to the (thin-walled tube) product forms that are typically used in the manufacture of endovascular stents. In the current work, we document the basic fatigue-crack growth properties of flattened thin-walled ( approximately 400 microm thick) Nitinol tubing in a 37 degrees C air environment. Crack-growth behavior is characterized over a wide range of growth rates ( approximately 6 orders of magnitude) and load ratios, that is, as a function of the alternating and maximum stress intensities, at 50 Hz. Limited experiments at both 5 and 50 Hz were also performed in 37 degrees C air and simulated body fluid to determine whether the cyclic frequency affects the fatigue behavior. Fatigue-crack growth-rate properties in such thin-walled Nitinol tube are found to be quite distinct from limited published data on other (mainly bulk) product forms of Nitinol, for example, bar and strip, both in terms of the relative fatigue thresholds and the variation in steady-state growth rates.

Alloys↗