Metal artifacts caused by gradient switching.
Abstract
In metal parts, e.g., implants or instruments, eddy currents can be induced from gradient switching if positioned off-center inside the MR scanner. For the first time, a systematic analysis of related artifacts was performed. Current strength increases in conjunction with increasing size of the part, increasing electrical conductivity, distance from isocenter, and increasing gradient strengths. A xy-plane oriented copper ring (d(o) = 20 mm, d(i) = 15 mm, 2 mm thick) was examined at isocenter and at x = 15 cm, y = z = 0. Comparisons of xy-, xz-, and yz-slices, recorded for both possibilities to select encoding directions, revealed effects from ramp-down of the slice-selection and ramp-up of the read-out gradient. Near the metal part, temporary inhomogeneities were superimposed to the static field and spin-dephasing signal loss resulted, despite using spin-echo technique. Artifacts depended on excitation and read-out bandwidth. For an equivalent titanium ring, conductivity related effects could not be ascertained but distinct susceptibility effects occurred. MR compatibility of implants/instruments therefore requires both low susceptibility and low conductivity.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Hansjörg Graf, Günter Steidle, Petros Martirosian, Ulrike A Lauer, Fritz Schick. 2005. Metal artifacts caused by gradient switching.. https://doi.org/10.1002/mrm.20524