PubMed · 16130665
Flow model for coupled-column gas chromatography systems.
Abstract
One of the challenges in performing comprehensive two-dimensional gas chromatographic separations is being able to predict the average linear velocities of the carrier gas in the two columns, especially when they have different diameters. The problem is compounded for loop-type modulators, where two thermal trapping zones that switch from hot to cold and back simultaneously are separated by a delay loop. If the linear velocity in the loop is not tuned to the length of the loop and the modulation period, the dual-stage modulation may cease to work properly. A model has been developed that calculates the flow rates in the columns and predicts appropriate delay loop dimensions for a given set-up. Additionally, the model determines the pressure ramp that needs to be used in order to maintain constant average linear velocity within the modulator loop throughout the course of the separation.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
James Harynuk, Tadeusz Górecki. 2005-09-09. Flow model for coupled-column gas chromatography systems.. https://doi.org/10.1016/j.chroma.2005.06.008