PubMed Health⌕ Search

PubMed · 16223254

Integrated microfluidic device for solid-phase extraction coupled to micellar electrokinetic chromatography separation.

Abstract

An integrated microdevice was utilized for the autonomous coupling of solid-phase extraction (SPE) to micellar electrokinetic chromatography (MEKC). Porous plugs of polymethacrylate polymer approximately 200 microm in length) were fabricated by ultraviolet irradiation in microchannels. Microcolumns of hydrophobic beads packed against the polymethacrylate plugs were utilized for the quantitative extraction of rhodamine B, yielding preconcentration factors over 200 for a 90-s extraction. The calculated detection limit for this dye was 60 fM. A sample of coumarin dyes were concentrated by SPE, eluted in a nonaqueous solvent from a separate on-chip reservoir, and injected by a gated valve onto a separate column for MEKC analysis. Using the integrated device, a completely automated sequence of extraction, elution, injection, separation, and detection were performed in less than 5 min. Observed separation efficiencies were high, with plate heights below 2 microm. The analysis was at least 3 times faster than semiautomated, conventional, solid-phase extraction, while requiring no user intervention. The design, fabrication, and autonomous operation of the device is discussed.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jeremy D Ramsey, Greg E Collins. 2005-10-15. Integrated microfluidic device for solid-phase extraction coupled to micellar electrokinetic chromatography separation.. https://doi.org/10.1021/ac0507789

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Recent progress in the use of ionic polymers as pseudostationary phases for EKC.

This review concerns the introduction, characterization, and application of polymeric pseudostationary phases (PSPs) for EKC since 2004. Achiral and chiral polymers and separations are reviewed, as is the application of polymeric PSPs for the combination of EKC with mass spectrometric detection.

Chromatography, Micellar Electrokinetic Capillary↗

MEKC: an update focusing on practical aspects.

This paper reviews recent methodological and instrumental advances in MEKC. Improvements in sensitivity arising from the use of on-line sample concentration (sweeping, stacking, and combination of both protocols) and derivatization (in-capillary reactions and coupling with flow-injection systems) and improvements in resolution obtained by changing the composition of the BGE (e.g., with organic modifiers, ionic liquids, nonionic and zwitterionic surfactants, mixed micelles, and vesicles) or using coated capillaries are discussed in detail. In addition, MS and LIF spectroscopy are examined in relation to their advantages and restrictions as applied to MEKC analysis. Some thoughts on potential future directions are also expressed.

Chromatography, Micellar Electrokinetic Capillary↗

Comparison of migration models for acidic solutes in micellar electrokinetic chromatography.

The validity of two models that explain the migration of ionisable solutes in micellar electrokinetic chromatography (MEKC), mobility model and retention factor model, has been tested. For this purpose, the mobility (mu) and retention factor (k) of a set of 10 phenolic compounds with different hydrophobicity and pKa values have been determined for several sodium dodecyl sulphate (SDS) concentrations and pH values, and fitted to the models. Results show that in general the retention factor model explains better the retention of ionisable solutes, although for hydrophilic compounds at low SDS concentration, mobility model can give better fits. The different drawbacks pointed out by several authors in relation to both models have been checked, and a deep evaluation of each one has been done. As a result we have observed that, while in the retention factor model the variation of k with pH and [SDS] always follows the same trend, the variation of mu with these variables mainly depends on the value of the binding constant of the neutral form of the solutes to the micelles, KHA(m), which plays a critical role in the fit of the mobility model. Also we provide rules and advices to set up the experimental conditions to apply each model to a particular solute.

Chromatography, Micellar Electrokinetic Capillary↗