PubMed · 16545960
Enhanced MALDI ionization efficiency at the metal-matrix interface: practical and mechanistic consequences of sample thickness and preparation method.
Abstract
Electrosprayed spots of varying thickness were evaluated for use as reproducible, homogenous, high efficiency MALDI samples. Thin samples on stainless steel plates were found to give exceptionally strong signals, as did the last layers of thick samples, when ablated down to the steel substrate. A small enhancement was also observed for thin samples on a gold substrate, and with a few-nanometer gold coating on top of a thick sample. Ion yields and intensity ratios can be understood in the context of the previously described quantitative MALDI model including the matrix-metal interfacial ionization potential reduction effect (Knochenmuss, R.; Anal. Chem. 2004, 76, 3179-3184). The absolute and relative stabilities of ion signals were found to be at least a factor of two better for the thin electrosprayed spots, compared to spots prepared by dried droplet methods.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Gregor McCombie, Richard Knochenmuss. 2006-03-20. Enhanced MALDI ionization efficiency at the metal-matrix interface: practical and mechanistic consequences of sample thickness and preparation method.. https://doi.org/10.1016/j.jasms.2006.02.005
Cite the original work for its findings. Save a collection to share your selection of sources.