PubMed · 16936647
Characterizing reflections from intraocular lens implants.
Abstract
OBJECTIVE: To develop a test method for characterizing glare from intraocular lenses (IOLs) and to confirm a clinical finding that the haptic insertion in the optic of a three-piece IOL produces extraneous line images. METHOD: The method consists of directing a collimated Gaussian laser beam to various parts of the IOL to be tested in a water-filled model eye. Reflected images produced in the retinal plane are photographed with a digital camera. RESULTS: A test method was developed to characterize the source of glare images from IOLs. The test method developed was used to confirm a clinical finding that the haptic insertion in the optic of a three-piece IOL produces extraneous line images. CONCLUSIONS: The method developed can be used to characterize and pin point the source of extraneous glare images from intraocular lens implants. The haptic insertion in the optic of a three-piece IOL has been identified as a source of line images.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R J Landry, I K Ilev, T J Pfefer, M Wolffe, J J Alpar. 2006-08-25. Characterizing reflections from intraocular lens implants.. https://doi.org/10.1038/sj.eye.6702540
Cite the original work for its findings. Save a collection to share your selection of sources.