Abstract
We report bright-field and dark-field surface-plasmon imaging using a modified solid immersion lens and a commercial objective of moderate NA in the epi configuration. The contrast and resolution are extremely good, giving well-resolved images of protein monolayers both in air and in water. We also describe a two-part solid immersion lens that allows the sample to be moved without degrading the image quality in any observable way. The merits of the two-part lens are discussed and compared to commercially available microscope objectives. Finally, we introduce a simple Green's function model that illustrates the key features of both bright-field and dark-field surface-plasmon imaging.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Jing Zhang, Mark C Pitter, Shugang Liu, Chungwah See, Mike G Somekh. 2006-11-01. Surface-plasmon microscopy with a two-piece solid immersion lens: bright and dark fields.. https://doi.org/10.1364/ao.45.007977