A ventilator generating a positive or negative internal compliance.

Abstract

This paper describes technical details of a ventilator for altering the resistive and elastic load placed on respiratory muscles during spontaneous breathing in intubated infants. Positive or negative values for ventilator resistance and/or ventilator compliance can be chosen by superimposing the weighted sum of the flow and/or the volume signal over the input to a pressure controller within the pressure feedback control system of the ventilator. The aim of the study was to compare values of the ventilator's compliance (Cv), as measured with a ventilation mechanics calculator, with those Cv values set by the ventilator's Cv control knob on the front panel. Another aim was to compare measured values of total compliance of a combined ventilator-lung model system (Ct) with the values expected according to theory where 1/Ct = 1/Cv + 1/C1m (Eq. alpha; C1m is the lung model's compliance). The Cv values set on the front panel were nearly identical to those measured (Cvm = 0.97 * Cvs + 0.54) over the whole tested range from -20 to +20 ml/kPa. Similarly, the measured Ct values were almost equal to those expected according to Eq. alpha; the standard deviation of the relative residuals was 2.7% for elastic loading and 12.4% for elastic unloading. We conclude that the ventilator described in this study can effectively provide both elastic loading and elastic unloading of spontaneous breathing, as expected according to theory.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

P Schaller, A Schulze. 1991. A ventilator generating a positive or negative internal compliance.. https://doi.org/10.3109/03009739109179274

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Optimal design of non-Newtonian, micro-scale viscous pumps for biomedical devices.

The present paper addresses the numerical optimization of geometrical parameters of non-Newtonian micro-scale viscous pumps for biomedical devices. The objective is to maximize the mass flow rate per unit of shaft power consumed by the rotor when an external pressure load is applied along the channel that houses the rotor. Two geometric parameters are considered in the optimization process: (i) the height of the channel that houses the rotor (H) and (ii), the eccentricity (epsilon) of the rotor. Three different micro-scale viscous pump configurations were tested: a straight-housed pump (I-shaped housing) and two curved housed pumps (L- and U-shaped housings). The stress-strain constitutive law is modeled by a power-law relation. The results show that the geometric optimization of micro-scale viscous pumps is critical since the mass flow rate propelled by the rotor is highly dependent on epsilon and H. Numerical simulations indicate that mass flow rate is maximized when epsilon approximately 0, namely when the rotor is placed at a distance of 0.05 radii from the lower wall. The results also show that micro-scale viscous pumps with curved housing provide higher mass flow rate per unit of shaft power consumed when compared with straight-housed pumps. The results are presented in terms optimized dimensions of all three configurations (i.e., H(opt) and epsilon(opt)) and for values of the power-law index varying between 0.5 (shear thinning fluids) and 1.5 (shear-thickening fluids).

Biomedical Engineering↗