PubMed · 1892351
Analytical goals for interference.
Abstract
Objective analytical goals for interference in clinical biochemical methods have not yet been advocated. We propose that, since total analytical error is ideally less than half the within-subject biological coefficient of variation (CVI), the maximum allowable systematic error produced by an interferent (I) is: I less than CVI-(1.96 CVA + SE), where CVA is the relevant experimental analytical imprecision and SE is the systematic error. Such a goal may be applicable also to non-specificity, matrix effects and carryover.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
X Fuentes-Arderiu, C G Fraser. 1991. Analytical goals for interference.. https://doi.org/10.1177/000456329102800414