[Oral temperature measurement using Craftemp].
Abstract
The performance of oral, electronically measured body temperature with Craftemp (Astra Meditec) in routine use in medical wards was investigated. Rectal measurement with conventional Hg-thermometer was used as a reference method. The Craftemp method turned out to be unacceptably inaccurate: 30.8 per cent of the measurements of rectal-oral temperature difference were subject to a variation around the mean larger than +/- 0.4 degree C. The standard deviation of the rectal-oral temperature differences was 0.51 degree C, and the range -1.2 degrees C to +3.0 degrees C. Until an alternative method for measurement of body temperature is proven accurate in the clinical routine, the established method of rectal measurement should not be abandoned.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
T Steen. 1990. [Oral temperature measurement using Craftemp].. https://pubmed.ncbi.nlm.nih.gov/2235474/