Apparent pacemaker malfunction due to ventricular blanking.

Abstract

Electrocardiographic interpretation in patients with dual chamber pacemakers requires an intimate understanding of their manner of functioning. This report describes a patient whose pacemaker was suspected of malfunction because of failure to sense premature beats. This sense-failure resulted from the fortuitous concurrence of the sensed portion of the premature beat with the ventricular blanking period and was resolved by reducing the low rate limit of the pacemaker. Three principles must be recognized in interpreting suspected failure of sensing: 1. The channel of origin of the pacing artifact must be identified. 2. The surface QRS morphology does not necessarily correspond to the signal detected by the pacemaker. 3. Failure to sense does not necessarily equal pacemaker malfunction.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

N D Berman, C A George, G B Duxbury. 1987. Apparent pacemaker malfunction due to ventricular blanking.. https://pubmed.ncbi.nlm.nih.gov/2436730/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

[CME-ECG 2].

Explore the source record for details and available documents.

Cardiac Complexes, Premature↗

Extrasystoles: side effect of kangaroo care?

OBJECTIVE: To present an unpublished reason for an arrhythmic electrocardiogram (ECG) recording during kangaroo care in a preterm infant. DESIGN: Case report. PATIENT: Preterm infant. MEASUREMENTS AND MAIN RESULTS: A preterm infant exhibited cardiac arrhythmia on the ECG monitor during kangaroo care, leading to interruption of kangarooing. Arrhythmia disappeared after placing the baby back into the incubator. The most likely reasons for arrhythmia were excluded. However, arrhythmia reappeared upon continuation of kangaroo care. ECG monitoring revealed the reason for the monitoring error. CONCLUSIONS: ECG monitoring during kangaroo care should cause error because of superimposed electric activity from the parent. Oxygen saturation represents a more reliable method of monitoring during kangaroo care.

Cardiac Complexes, Premature↗