Are ECG Welsh cup electrodes effectively cleaned?
Abstract
Re-usable Welsh electrocardiograph (ECG)-electrodes are potential vehicles for cross infection. This study confirmed that in-use ECG-electrodes are frequently contaminated with organisms such as coagulase-negative staphylococci and Gram-negative bacilli. The efficacy of five cleaning procedures was evaluated. Immersing the electrodes in water at 60 degrees C for one hour was the most effective method of decontamination tested, following challenge with a standardised suspension of Staphylococcus saprophyticus. Significant contamination persisted following simple cleaning measures. It is suggested that this was promoted by the inadequate removal of electrode gel which provided a protective environment for microorganisms.