PubMed Health⌕ Search

PubMed · 2580281

Dual chamber pacemakers: upper rate behavior.

Abstract

Upper rate management of a dual chamber pacemaker requires sensing of the atrium and limitation of ventricular response to desired levels. Limitation of ventricular rate response is accomplished by setting atrial channel refractoriness which consists of two separate and continuous intervals, the atrioventricular interval (AV interval) and the atrial refractory interval, after the ventricular pace or sense event (AR interval). A P-wave that falls within the total atrial refractory interval (TARI) remains unsensed and one that falls beyond the TARI is sensed. If the upper rate limit interval (URI) is programmed to equal the TARI, the upper rate limit will occur by development of sudden AV block. If the URI is of greater duration (lower allowed rate) than the TARI, the difference in interval between the two is the Wenckebach interval (WI) and is the duration of a response plateau, when the atrial coupling interval is less than the upper rate interval. All dual chamber timing cycles can be interpreted in terms of atrial refractoriness and upper rate limitation with consideration of the lower rate interval, and the ventricular refractory and blanking intervals.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

S Furman. 1985. Dual chamber pacemakers: upper rate behavior.. https://doi.org/10.1111/j.1540-8159.1985.tb05751.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Sarcoidosis presenting with tachy- and brady-arrhythmias.

We describe a case report of a 49-year-old man admitted to a cardiology unit with a series of arrhythmias with no initially obvious aetiology. Further assessment and the use of cardiac magnetic resonance imaging and histology allowed a diagnosis of cardiac sarcoidosis to be made. Cardiac sarcoidosis is a major cause of death in patients with systemic sarcoidosis. Cardiac magnetic resonance imaging is an additional diagnostic tool for this condition without ionizing radiation exposure.

Bradycardia↗

Non-sinus bradyarrhythmias in very low birth weight infants.

Transient bradycardias occur frequently in preterm infants with apnea of prematurity. However, occurrence of non-sinus bradyarrhythmias in very low birth weight (VLBW) infants has received very little attention. This case series documents transient bradyarrhythmias in VLBW infants. Most frequently observed arrhythmia was a non-sinus atrial rhythm followed by junctional escape rhythm. Ventricular rhythm with wide QRS complexes occurred rarely. The majority of these episodes occurred in the absence of oxygen desaturation. Increase in vagal tone is presumed to mediate this response.

Bradycardia↗

QT interval correction in patients with cirrhosis.

INTRODUCTION: QT interval prolongation is a common electrophysiological abnormality in patients with cirrhosis. As QT interval varies with the heart rate, many QT correction formulas have been proposed, the Bazett's one being the most criticized because it over-corrects the QT interval and may be misleading. This study focused on the QT-RR relationship in patients with cirrhosis to derive a population-specific QT correction formula. METHODS: One hundred cirrhotic patients of different etiology and severity and 53 healthy controls comparable for age and sex were enrolled. The QT-RR relationship was analyzed in patients by five regression analysis models to derive the population-specific QT-RR equation. The QTc was calculated and compared with those calculated by four common QT correction formulas (Bazett, Fridericia, Framingham, and Hodges). The correlation coefficient QTc-RR was calculated as a measure of the independence of QTc from the original RR interval. RESULTS: In patients the QT-RR relationship was best described by the power equation "QT = 453.65 x RR1/3.02" (R2 = 0.41), similar to the Fridericia's formula. Bazett's formula led to the longest QTc (P < 0.0001), which was still significantly influenced by the RR interval (R = -0.39; P < 0.0001), while the estimated equation led to a QTc value not influenced by RR (R = -0.014). CONCLUSION: Bazett's correction should be avoided in patients with cirrhosis because it still provides a rate-dependent QTc value and might be misleading, particularly when assessing the overall preoperative cardiac risk and the effect of drugs affecting the QT interval. In its place, our formula or that of Fridericia can be confidently employed.

Bradycardia↗