PubMed · 2598580
A microprocessor temperature-controlled air delivery system for dentinal hypersensitivity testing.
Abstract
A new microprocessor temperature controlled air delivery system for determining cold and warm temperature thresholds of dentinal hypersensitivity has been designed and built. The temperature range of the unit is -5 degrees to + 85 degrees C. Air is delivered from a typical dental unit air syringe nozzle at a relatively gentle flow rate of 0.2 cu ft/min at operator selected temperatures, with an accuracy of +/- 0.2 degrees C. Within seconds, 1 degree to 5 degrees temperature increments or decrements are accomplished. Larger, i.e., 20 degree increments or decrements, require approximately 15 seconds. The unit can operate from standard 110 V/60 Hz or 230 V/50 Hz outlets and requires a source of compressed air at 60 psig, equivalent to that provided by dental operatory-type air compressors. This report provides a description of the unit, its operation, and data from two clinical studies in which it was used successfully to determine temperature thresholds of dentinal hypersensitivity.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
P Person, E E Demand, L Koltun, L M Spindel. A microprocessor temperature-controlled air delivery system for dentinal hypersensitivity testing.. https://pubmed.ncbi.nlm.nih.gov/2598580/