Abstract
A novel vertebral prosthesis is presented. The prosthesis was developed for surgical procedures requiring the resection of a complete vertebral body and the adjacent intervertebral discs, the design objective being to develop a flexible implant that would be robust enough to withstand the in vivo stress environment of the human spine. In theory, a flexible implant should preserve a more normal range of motion and apply less stress to surrounding tissue than a rigid implant. A prototype implant was constructed so as to combine a rigid stainless steel structure with flexible silicon rubber elements in order to form an implant with static and dynamic mechanical characteristics similar to those of the anterior spinal column. Implant flexibility characteristics were determined from ex vivo stress-strain behaviour during bending and compressive creep testing. Results from the bending tests indicated good agreement for the lateral and sagittal bending characteristics in comparison with in vitro bending tests of human lumbar motion segments. Comparison of the implant compressive creep response with similar in vitro tests on human lumbar intervertebral discs also demonstrated similarities in the time-dependent mechanical parameters.