A monolithic patch-clamping amplifier with capacitive feedback.

Abstract

Patch-clamping is an established method for directly measuring ionic transport through cellular membranes with sufficient resolution to observe open/close transitions of individual channel molecules. This paper describes an alternative technique for patch-clamping which uses a capacitor as the transimpedance element. This approach eliminates bandwidth and saturation limitations experienced with resistive patch-clamping amplifiers. A complete monolithic design featuring an on-chip operational amplifier, a capacitor array with gain-ranging from 30 pF down to 0.03 pF, and reset and gain ranging switches has been fabricated using 5 microns CMOS technology. It is shown that the voltage noise of the CMOS operational amplifier limits the overall noise performance, but that performance competitive with conventional instruments can be achieved over a 10 kHz bandwidth, at least for small input capacitances (less than or equal to 5 pF). Results are presented along with an analysis and comparison of noise performance using both resistive and capacitive elements.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J Prakash, J J Paulos, D N Jensen. 1989. A monolithic patch-clamping amplifier with capacitive feedback.. https://doi.org/10.1016/0165-0270(89)90099-x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Design and evaluation of shielded gradient coils.

Techniques are described for the design of shielded gradient coils for superconducting MRI systems. These design methods are suited for constructing the most efficient gradient coil that meets a specified homogeneity requirement. Tradeoffs in coil design of efficiency with coil size and gap size are discussed. Residual eddy currents from coils constructed with a finite number of wires are calculated and give guidelines for the construction of efficient, whole-body gradient coils.

Amplifiers, Electronic

A piezoelectric respiratory monitor for in vivo NMR.

This report describes a simple electronic device employing a piezoelectric element which serves as a sensitive detector of motion. The device is useful as a monitor of respiratory motion for nuclear magnetic resonance animal experiments in vivo. It can also provide a trigger pulse for respiratory gating experiments.

Amplifiers, Electronic

Correction for liquid junction potentials in patch clamp experiments.

This chapter describes corrections that have to be applied to measured membrane potentials in patch clamp experiments. Some of them [Eqs. (1)-(3)] are required regardless of the nature of the reference electrode (in the Ringer's solution bath) whenever the pipette-filling solution is different from the bath solution. They represent the liquid junction potentials that are present at the pipette tip before patch formation. In addition, corrections have to be applied when the bath solution is being changed during a measurement (i.e., after seal formation). In that case the following rules apply. (1) The new solution should never get into contact with the bare silver/silver chloride wire of the reference electrode. This requirement is best met by using a salt bridge. (2) The "best" salt bridge is a 3 M KCl bridge with an abrupt KCl-bath fluid boundary at its tip (see above). This bridge does not require any additional potential corrections, but it may lead to KCl poisoning of the bath or become contaminated by solutions used previously. (3) Local solution changes (microperfusion by puffer pipette, U tool or sewer pipe arrangements) as well as recessed KCl bridges require additional corrections, which (together with the simple liquid junction potential correction) are approximately given by Eqs. (6)-(8). It should be stressed that all equations given here represent approximate corrections, since liquid junction potentials are thermodynamically ill-defined. This is particularly relevant for Eqs. (6) and (7) where the sum of two liquid junction potentials appears.

Amplifiers, Electronic