PubMed HealthSearch

PubMed · 2718054

Complications with the variable spinal plating system.

Abstract

From January 1986 to June of 1987, 40 patients underwent transpedicle fixation and fusion using the variable spinal plate system. Nineteen patients had undergone surgery at the same level or levels, and 21 patients had undergone no previous surgery. Diagnostic categories include spondylolisthesis, thoracolumbar and lumbar fractures, internal disc derangement, spinal stenosis, pseudarthrosis, mechanical instability, and fracture mal-union. Minimum follow-up has been 14 months, with the average being 20 months. Overall results showed 13 excellent, 12 good, seven fair, and eight poor. The overall complication rate was 45%. In those patients undergoing no previous surgery, it was 29%, but with those patients having previous surgery, it was 63%. Most of these complications were minor in nature and resolved before discharge. Implant failure occurred in seven patients, and consisted of screw breakage. Design modifications currently available should help minimize this complication. Although this method of internal fixation and fusion is technically demanding and has a high complication rate, it is considered to be indicated in lumbar fractures, revision of pseudarthrosis, spondylolisthesis with or without reduction, and failed surgery with marked instability.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

T S Whitecloud, J C Butler, J L Cohen, P D Candelora. 1989. Complications with the variable spinal plating system.. https://doi.org/10.1097/00007632-198904000-00027

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Fixation of fractures by twisted plates. A new concept of osteosynthesis.

A new twisted plate has been designed for biplane fixation of diaphyseal fractures of long bones. Comparative mechanical testing of the new twisted plate and conventional flat plates showed that the twisted plate was 40% stronger when a bending moment was applied and 132% stronger to a twisting moment. The overall improvement in strength was 90.5%. Clinical trials have been carried out and this paper describes the use of the new plate.

Bone Plates