PubMed · 2921973
Dose delivery error detection by a computer-controlled linear accelerator.
Abstract
A commercial dual photon energy, computer-controlled linear accelerator has a complex collimation and beam delivery system. For accurate dose delivery, six separate motorized elements must be properly positioned for a given beam selection. Experimentally instituted misalignments of the primary electron scattering foils, the primary collimator, the flattening filter, the scattering foil carousel, and the backscatter shield or shutter produced significant dose delivery errors. The ability of the Mylar window monitor chamber to detect these errors was examined for x-ray beams. The fault detection system failed to interrupt dose delivery in a number of situations where the error in dose per monitor unit delivered ranged from -6% to +270% of the calibrated value.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
M Loyd, H Chow, J Laxton, I Rosen, R Lane. Dose delivery error detection by a computer-controlled linear accelerator.. https://doi.org/10.1118/1.596384
Cite the original work for its findings. Save a collection to share your selection of sources.