Spatial distribution of dose in computed tomography with special reference to thin-slice techniques.

Abstract

The spatial dose distribution in a cylindrical polystyrene phantom with a diameter of 200 mm was measured for seven computed tomography (CT) scanners. The measurements were performed in the head mode and mainly for narrow slices in the range 1.5 to 4 mm. Both radial and axial dose profiles were measured and the dose distribution for multiple-scan procedures was calculated. The ratio between the surface and centre doses for a single scan varied between the extremes of 1.8 and 4.3 and was generally higher for narrow than for wide slices. With multiple nominally contiguous scans the difference in absorbed dose between surface and centre locations in the object decreased, on account of scattered radiation. The CT dose index for centre locations varied considerably between the tested scanners, with a range from 5.6 to 27.2 mGy per nominal 100 mAs. For a simulated multiple-scan procedure, comparable to a CT examination of the orbits, the multiple-scan average dose varied between 4.3 and 16.4 mGy per nominal 100 mAs.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

U Moström, C Ytterbergh. Spatial distribution of dose in computed tomography with special reference to thin-slice techniques.. https://pubmed.ncbi.nlm.nih.gov/2962618/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Patient dose measurements in diagnostic radiology procedures in Korea.

This study is the first nationwide investigation aimed at estimating the patient dose for radiographic examinations in Korea including gastrointestinal studies, computed tomography and mammography. The survey data from 161 hospitals and the dose data from 32 hospitals were analysed. The third quartile entrance surface dose, dose area product (DAP), weighted CT dose index (CTDIw) and mean glandular dose (MGD) were reported. All the estimated doses were less than the stated International Atomic Energy Agency (IAEA) reference levels for radiographic examinations. However, DAPs for the fluoroscopic examinations had higher dose values than the IAEA reference levels. In addition, the CTDIw and MGD were lower than the IAEA reference levels.

Body Burden↗

The study of new calibration features in the Harshaw TLD system.

The Harshaw TLD system has three key calibration procedures: the Reader, the Dosemeter and the Algorithm. These functions must be properly calibrated for the system to achieve the optimum results. For the conventional reader and dosemeter calibration, Harshaw TLD recommends a pre-fade and a post-fade of 24-48 h when calibrating the system for LiF:Mg,Ti type dosemeter. It is also recommended that keeping the fade time consistent is important to maintain the quality of the system performance. In recent years, new calibration features have been introduced into the Harshaw TLD models 6600 and 8800 operating systems. These new features are Auto Calibration, Auto QC and Auto Blank, and they give the user the ability to set up the clear-expose-read process to be performed automatically in a sequence for each dosemeter. This saves processing time and keeps the fade time the same. However, since the fade time is near zero, will it affect the TLD system calibration factors? What should the user expect? This paper presents a study of the effect of Auto Calibration/Auto QC to the TLD operation.

Body Burden↗

Proton and photon absorbed-dose conversion coefficients for embryo and foetus from top-down irradiation geometry.

Absorbed-dose conversion coefficients are calculated for the embryo of 8 weeks and the foetus of 3, 6 or 9 months when the mother's body is exposed to protons and photons from top-down (TOP) direction. It provides data sets in addition to other standard irradiation geometries published previously. The TOP-irradiation geometry is considered here, because high-energy particles are often peaked from the TOP direction onboard aircrafts. The results show that absorbed-doses from high-energy particles could be underestimated significantly if isotropic (ISO) irradiation geometry is assumed. For protons of 100 GeV, absorbed-doses from TOP irradiation are approximately 2.3-2.9 times higher than the doses from ISO irradiation for different foetal ages. For 10 GeV photons, foetal doses from TOP irradiation are approximately 6.8-12 times higher than the doses from ISO irradiation. The coefficients from TOP-irradiation geometry are given in wide energy ranges, from 100 MeV to 100 GeV for protons and from 50 keV to 10 GeV for photons. They can, therefore, be used in various applications whenever exposure from the TOP-irradiation direction is concerned.

Body Burden↗