Computer simulation of image intensifier-based computed tomography detector: vascular application.

Abstract

This present study reports the results of a computer simulation whose aim was to predict the low-contrast imaging performance of which a conventional x-ray image intensifier with charge coupled device (CCD) camera would be capable if incorporated into a computed tomography (CT) volume imager. A vascular imaging task was modeled in our simulation. The effects of detector noise, x-ray exposure levels, analog-to-digital conversion (ADC) precision and residual levels of detected x-ray scatter were considered. The results of this simulation indicate that the low-contrast imaging performance of an image intensifier-based CT system was most limited by the CCD detector readout noise. Given this limitation the detection of greater than about 100,000 detected photons/pixel/projection gave marginal improvement in low-contrast resolution. At these exposures 12 bit ADC precision resulted in little additional image noise. The effects of detecting scattered x rays are twofold; decreasing the signal-to-noise ratio associated with our modeled artery and introducing a cupping artifact. Based on the results from the simulation, it appears that an image intensifier-based CT system is a feasible concept from a noise viewpoint, if the anticipated imaging task is intravenous angiography.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R Ning, R A Kruger. Computer simulation of image intensifier-based computed tomography detector: vascular application.. https://doi.org/10.1118/1.596250

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Transhepatic vascular access for diagnostic and interventional procedures: techniques, outcome, and complications.

Early or multiple cardiac catheterizations may result in occlusion of the femoral veins prohibiting their use. The internal jugular or sub-clavian approach may be an appropriate alternative. However, these approaches may not be suitable in patients with surgical interruption of the superior vena cava. In other patients, they may not allow easy access to certain areas of the heart. The transhepatic approach is an important alternative route for performing cardiac catheterization and interventions. Depending on the planned procedure, it may be the preferred route to perform the cardiac catheterization even in the presence of patent femoral veins. The indications, technical details and potential complications of this approach are discussed.

Angiography↗

Embolization for a bleeding pelvic fracture in a patient with persistent sciatic artery.

Emergency embolization for a bleeding pelvic fracture was performed in a patient with persistent sciatic artery, a rare anatomic variation in which the internal iliac artery continues to the popliteal artery. Successful hemostasis was achieved without complications by means of selective catheterization into each branch of the internal iliac artery. Awareness of this unusual vascular anatomy is critically important to avoid serious complication of emergency embolization.

Angiography↗