PubMed · 3784990
A new pulse sequence for determining T1 and T2 simultaneously.
Abstract
Determination of the relaxation times T1 and T2 which are important for tissue characterization generally requires the use of different pulse sequences in magnetic resonance imaging. In this study, a new pulse sequence which facilitates simultaneous determination of the T1 and T2 times is presented. Determination takes place in this case pixel by pixel from the measured images. The measuring time corresponds in this case approximately to that of a normal spin-echo sequence with long repetition time and two data acquisitions. The functional dependence of the accuracy of the T1 and T2 determination upon external errors, e.g., angle of rotation errors, is discussed. The tissue contrast behavior of the individual echoes is shown and its dependence on pulse parameters is explained.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R Graumann, H Fischer, A Oppelt. A new pulse sequence for determining T1 and T2 simultaneously.. https://doi.org/10.1118/1.595867
Cite the original work for its findings. Save a collection to share your selection of sources.