PubMed · 4060832
[Unusual rhythm disorders with a DDD pacemaker].
Abstract
A tachycardia caused by an inadvertent change of the lead connections in a patient with a DDD pacemaker is reported. This was demonstrated by operating in different modes (i.e. AOO, VOO) by displaying the marker channel graphically and by X-ray examination revealing unchanged lead positions.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K H Konz, K D Schick, R Clausnitzer, C Huth, L Seipel. 1985. [Unusual rhythm disorders with a DDD pacemaker].. https://pubmed.ncbi.nlm.nih.gov/4060832/