PubMed · 6200856
Atrial trigeminy induced by a normally functioning DVI pacemaker.
Abstract
Atrial trigeminy was seen in a patient with a normally functioning DVI pacemaker (Byrel-5992). Atrial competition resulted from coincidence of sinus and atrial escape rates together with late ventricular sensing. This resulted in atrial output stimuli occurring within the QRS complex. Alternate impulses occurred late enough to conduct to the ventricle after recovery of the AV node. Reprogramming the pacemaker to either a faster DVI rate or to VVI abolished the arrhythmia.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
R N Millar, I W Obel. 1984. Atrial trigeminy induced by a normally functioning DVI pacemaker.. https://doi.org/10.1111/j.1540-8159.1984.tb04898.x
Cite the original work for its findings. Save a collection to share your selection of sources.