PubMed · 6429096
Pion treatment procedures and verification techniques.
Abstract
Procedures and techniques developed for the negative pi-meson (pion) radiotherapy program at the Los Alamos Meson Physics Facility, Los Alamos, NM, are reviewed and described. A particular pion patient is followed through the entire planning and treatment sequence to describe CT scanning procedures, bolus and collimator and treatment techniques developed to minimize positioning errors (less than 5 mm). Comparison of 2-D and 3-D isodose calculations developed at Los Alamos showed differences of less than 10% attributable to multiple scattering effects and the computational models used. Treatment verification methods using in vivo ion chamber dosimetry generally confirmed the prescribed dose delivery within 10% and using TLD within 18%.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S R Zink, S E Bush, C J Gilman, R H Hilko, R K Justice, E C Osborne, A R Smith, P A Berardo. 1984. Pion treatment procedures and verification techniques.. https://doi.org/10.1016/0360-3016(84)90303-1
Cite the original work for its findings. Save a collection to share your selection of sources.