Abstract
A cold stage has been constructed for the Philips EM300 (and EM301) electron microscopes for investigating the structure of frozen-hydrated biological specimens. The stage entails minimal alterations to the instrument and is capable of a resolution better than 10 A at the normal operating temperature of -120 degrees C. Frozen specimens can be readily exchanged without condensation or warming up, and maintained in the stage over periods of several hours without detectable deterioration.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
K A Taylor, R A Milligan, C Raeburn, P N Unwin. 1984. A cold stage for the Philips EM300 electron microscope.. https://doi.org/10.1016/0304-3991(84)90197-9