An automated simultaneous transmural cardiac mapping system.

Abstract

The origin and propagation sequence of cardiac depolarization in situ requires accurate simultaneous three-dimensional information from multiple sites. Likewise, the mechanism underlying an arrhythmia can often be elucidated by determining the course of impulse propagation through the heart, particularly if information can be obtained from multiple sites simultaneously, allowing analysis of transient or rapidly occurring events. The most significant problem with obtaining such detailed continuous information is the large amount of data storage required as well as the need for rapid analysis. In the present system these problems are overcome by immediate conversion of all electrograms from analog to digital for all subsequent storage and processing. The bipolar electrogram information is acquired from 240 cardiac sites simultaneously at a sampling rate of 2 kHz with continuous and total data storage of up to 60 min. Rapid two-dimensional isochronic maps at multiple depths (effective 3-dimensional information) are presented via computer-generated interactive graphics. System design permits easy expansion to almost 2,000 simultaneous sites. Surgical electrophysiological intraoperative studies in humans are performed at an operating room located 2,000 ft from the computer facility with all communications carried by a fiber-optic link. The system allows both experimental and clinical cardiac mapping from multiple sites from a single cardiac depolarization, minimal redundancy of costly hardware, and direct rapid visualization of all original electrogram data.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

F X Witkowski, P B Corr. 1984. An automated simultaneous transmural cardiac mapping system.. https://doi.org/10.1152/ajpheart.1984.247.4.h661

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Safety of nerve conduction studies in patients with implanted cardiac devices.

Patients with implanted cardiac devices and their physicians may defer important electrodiagnostic testing because of anxiety about potential negative effects on the device. To determine the safety of routine nerve conduction studies (NCS) in this population, 10 patients with permanent dual-chamber pacemakers of various types and five patients with implanted cardiac defibrillators (ICD) underwent nerve stimulation at sites commonly used during NCS. The implanted cardiac device was interrogated before and after the study and there was continuous monitoring of the surface electrocardiogram (ECG) and atrial and ventricular electrograms. Electrical impulses generated during routine NCS were never detected by the sensing amplifier and did not affect the programmed settings of the implanted cardiac device. We conclude that routine NCS is safe in patients with implanted cardiac pacemakers with bipolar sensing configurations and defibrillators.

Arrhythmias, Cardiac↗

[Pulmonary diseases and heart function].

The most severe cardiac sequel to lung disease is the load on the right ventricle due to pulmonary hypertension with the development of a cor pulmonale. This is characterized by hypertrophy and/or dilatation of the right ventricle because of a primary impairment of lung function and/or lung structure. The most important pathomechanisms for the development of pulmonary hypertension are vessel obliteration, mechanical lesions, primary vascular or extra-vascular inflammation and hypoxic vasoconstriction. Chronic obstructive pulmonary disease (COPD) is one of the most important reasons for chronic cor pulmonale. A further very common reason is obstructive sleep apnea syndrome, especially if combined with a COPD. In this case, the prevalence of cor pulmonale can reach 80%. The development of a chronic cor pulmonale is the most striking negative prognostic factor for these patients. Only 30% of COPD patients with cor pulmonale survive longer than 5 years, and only early detection of the disturbances to respiration which might potentially lead to cor pulmonale and their subsequent therapy are able to improve the patient's prognosis. Furthermore, pulmonary diseases may also have an impact on the left heart side in terms of an impairment of left heart function or by inducing severe arrhythmias . Thus, lung diseases may have both a significant impact on right and left heart performance.

Arrhythmias, Cardiac↗