Automated external defibrillation: laboratory evaluation.

Abstract

Twelve samples of ventricular fibrillation were fed into nine automated external defibrillator-pacemakers ( AEDP , "Heart Aid") of recent design. All the devices recognised and defibrillated ventricular fibrillation in seven of the samples within 30 sec. None of the devices reacted to two of the samples; in the remaining three there was inter-device variation ranging from an appropriate response to no response, as well as inappropriate pacing or delay in recognition and treatment. Poor recognition of some ventricular fibrillation waveforms with considerable inter-device variation limits the usefulness of this model. A new prototype responded more consistently and future models may be of value in community resuscitation. The difficulty of evaluating the diagnostic capability of AEDP devices in clinical use makes comprehensive laboratory testing essential prior to release.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

E Rosenthal, D Carroll, R Vincent, D A Chamberlain. 1984. Automated external defibrillation: laboratory evaluation.. https://doi.org/10.1016/0167-5273(84)90079-2

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A biventricular ICD system with biventricular defibrillation.

We describe the case of a 59-year-old gentleman with severe dilated cardiomyopathy requiring implantation of a dual-chamber biventricular implantable cardioverter-defibrillator (ICD). High defibrillation thresholds (DFT) were encountered at implant with an inadequate defibrillation safety margin. Testing of all possible shock vectors/polarities with and without the SVC coil and optimization of the distal RV coil position all proved inadequate. A satisfactory defibrillation safety margin was achieved following placement of a second lead in the coronary sinus to enable biventricular defibrillation. This case highlights an additional strategy for combating high DFTs and is an option even in dual-chamber biventricular ICD systems.

Electric Countershock↗