PubMed · 6818217
A new system for ventilating with high-frequency oscillation.
Abstract
We describe simple high-frequency oscillation systems that incorporate a CO2 absorber and supply O2 on a need basis. These systems have the advantage of easy control of mean airway pressure and airway hydration and negligible loss of oscillatory tidal volume. Experiments done at constant tidal volume showed that as frequency (and hence total ventilation) increased, arterial CO2 tension (PaCO2) decreased. The fall in PaCO2 occurred until frequency reached approximately 20 Hz; above 20 Hz further increases in frequency had little or no effect on PaCO2. Because of their practical advantages the techniques described here may be quite useful in a clinical setting where an oscillator, rather than jet-type high-frequency, ventilation system is desired.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Y K Ngeow, W Mitzner. 1982. A new system for ventilating with high-frequency oscillation.. https://doi.org/10.1152/jappl.1982.53.6.1638. https://pubmed.health/record/pubmed%3A6818217/a-new-system-for-ventilating-with-high-frequency-oscillation
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.