Optimal short scan convolution reconstruction for fanbeam CT.

Abstract

The problem of using a divergent fan beam convolution reconstruction algorithm in conjunction with a minimal complete (180 degrees plus the fan angle) data set is reviewed. It is shown that by proper weighting of the initial data set, image quality essentially equivalent to the quality of reconstructions from 360 degrees data sets is obtained. The constraints on the weights are that the sum of the two weights corresponding to the same line-integral must equal one, in regions of no data the weights must equal zero, and the weights themselves as well as the gradient of the weights must be continuous over the full 360 degrees. After weighting the initial data with weights that satisfy these constraints, image reconstruction can be conveniently achieved by using the standard (hardwired if available) convolver and backprojector of the specific scanner.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D L Parker. Optimal short scan convolution reconstruction for fanbeam CT.. https://doi.org/10.1118/1.595078

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Longitudinal image deblurring in spiral CT.

PURPOSE: To assess the feasibility of digital deconvolution techniques to improve longitudinal resolution of spiral computed tomography (CT) multiplanar reformations and evaluate how technical factors in deconvolution affect longitudinal resolution, noise, and edge ringing. MATERIALS AND METHODS: Longitudinal line spread function (LSF) of the system was estimated from longitudinal reformations of transaxial spiral CT images of a step test phantom. By using the estimated LSF, longitudinal reformations of the phantom and three clinical spiral CT studies were deconvolved by the methods of Wiener filtering and constrained iterative deconvolution. Edge ringing and image noise were quantified for Wiener filtering and constrained iterative deconvolution. RESULTS: Longitudinal reformations were substantially deblurred and resolution improved after deconvolution. Anatomic boundaries in clinical images were more clearly delineated after restoration. The methods of Wiener deconvolution and constrained iterative deconvolution improved the sharpness of the phantom step boundary at the expense of increased edge ringing and image noise. CONCLUSION: In longitudinal spiral CT reformations, blurring along the longitudinal axis can be reduced by Wiener filtering or constrained iterative deconvolution.

Technology, Radiologic↗