Coulometric microrespirometry.
Abstract
Capacitive coulometry is based on the automatic quantitative replacement of O2 consumed by an animal in a closed system with electrolytic O2, produced by discharging a capacitor through a CuSO4 solution. The sensitivity of the device is better than 0.1 nl. The unit volume of O2 produced (1 nl) is both accurate and precise within 1% in a temperature range from 2.8 to 40 degrees C. The upper limit of recordable O2 consumption is 1 ml/h. The microrespirometer can be autoclaved, making the technique ideally suited for metabolic studies in microbiology, cell and organ cultures, and comparative physiology.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
A A Heusner, J P Hurley, R Arbogast. 1982. Coulometric microrespirometry.. https://doi.org/10.1152/ajpregu.1982.243.1.r185