The geometric transfer function component for scintillation camera collimators with straight parallel holes.

Abstract

A theoretical approach has been developed that allows the geometric transfer function component for conventional scintillation camera collimators to be predicted in closed form. If transfer function analysis is to be useful in describing imaging system performance, the image of a point source must not depend on source position in a plane parallel to the detection plane. This shift invariance can be achieved by analysis of system response in terms of an effective point spread function, defined as the normalised image of a point source that would be obtained if the camera collimator were uniformly translated (but not rotated) during image formation. The geometric component of the corresponding effective transfer function is shown to be expressed simply by the absolute square of the two-dimensional Fourier transform of a collimator hole aperture, with the spatial frequency plane scaled by a factor which depends on collimator length, source-to-collimator distance, and collimator-to-detection plane distance. Closed form algebraic expressions of the geometric transfer function have been obtained for all four common hold shapes (circular, hexagonal, square and triangular). Monte Carlo simulations and experimental measurements have shown these theoretical expressions to be highly accurate.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

C E Metz, F B Atkins, R N Beck. 1980. The geometric transfer function component for scintillation camera collimators with straight parallel holes.. https://doi.org/10.1088/0031-9155%2F25%2F6%2F003

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Fast three dimensional magnetic resonance imaging.

To reduce the scan time in three-dimensional (3D) imaging, the authors consider alternative trajectories for traversing k-space. They differ from traditional 3D trajectories, such as 3DFT, in that they employ time-varying gradients allowing longer readouts and in turn a reduced scan time. Some of these trajectories reduce by an order of magnitude the number of excitations compared with 3DFT and provide flexibility for trading off signal-to-noise ratio for scan time. Other concerns are the minimum echo time and flow/motion properties. As examples, the authors show two applications: A 3D data set of the head (field of view of 30 x 30 x 7.5 cm and resolution of 1.5 x 1.5 x 1.5 mm) acquired in 56 s using a stack of spirals in 3D k-space; and a 3D movie of the heart (20 x 20 x 20 cm field of view, 2 x 2 x 2 mm resolution, and 16 time frames per cardiac cycle) acquired in 11 min using a cones trajectory.

Fourier Analysis

The application of Fourier deconvolution to reaction time data: a cautionary note.

The Fourier transform method in conjunction with frequency domain smoothing techniques has been suggested as a powerful tool for examining components in a serial, additive reaction time model (P. L. Smith, 1990). Robustness and sensitivity to violations of the assumptions of serial model of this method are evaluated. When an incorrect distribution was used in recovering an unobserved component, results gave no information to show that an incorrect distribution was used, and the results were just as interpretable as those obtained using the correct distribution. These results demonstrate that the assumptions underlying the method cannot be assessed by the result of deconvolution, and the method cannot show that the purported component is actually from the serial combination.

Fourier Analysis