PubMed HealthSearch

PubMed · 7386653

Dual-path capacitance compensation network for microelectrode recordings.

Abstract

Commercially available microelectrode preamplifiers have a built-in capacity compensation circuit. When a shielded cable is utilized to transmit the signal from the microelectrode to the preamplifier the compensative ability of this circuit may be inadequate and may result in a less than desired frequency response. We describe a dual-path capacitance compensation circuit that resolves this problem without impairment of the shielding properties.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R McGillivray, R Wald. 1980. Dual-path capacitance compensation network for microelectrode recordings.. https://doi.org/10.1152/ajpheart.1980.238.6.h930

Cite the original work for its findings. Save a collection to share your selection of sources.