Temperature field estimation using electrical impedance profiling methods. II. Experimental system description and phantom results.

Abstract

An electrical impedance tomography system has been developed and tested for the purpose of thermal imaging. Since impedance changes with temperature, images of impedance subtracted from normothermic baselines will provide a map of temperature data. A system was designed to be operational at 10-50 kHz and to utilize 16 external electrodes around the periphery of a tissue-equivalent phantom encompassing the region of interest. These electrodes serve as current sources for the 5 mA constant-current inputs and are also used for reading differential voltages. Hyperthermia treatments for cancer require that internal thermometry probes be inserted into the tumour volume. Linear arrays of electrodes with thermometry tracks for micro-dimension thermometry serve this function, as well as providing localized voltage measurements in the region of interest. The embedded temperature sensors provide a quality assurance and calibration standard for the linear arrays in reconstruction of impedance profiles. Results of transient heating experiments with conductive and ultrasound heating are shown where image reconstruction is performed using a finite element model. Temperature predictions in these studies were accurate to better than 1 degree C on average when using information from surface electrodes combined with internal linear arrays. Maximum temperature errors, however, was found to be > 5 degrees C which suggests that further noise reduction during data acquisition and improvements in the reconstructions algorithms are needed.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

M J Moskowitz, K D Paulsen, T P Ryan, D Pang. Temperature field estimation using electrical impedance profiling methods. II. Experimental system description and phantom results.. https://doi.org/10.3109/02656739409009345

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Gastric emptying patterns of a liquid meal in newborn infants measured by epigastric impedance.

Epigastric impedance was used to measure the gastric emptying patterns of a liquid non-caloric meal (5 mL water kg-1) in 30 healthy newborn infants. Twenty-six mature infants were examined in the first eight days of life, and four preterm infants were examined within 6 weeks after birth. The recordings consisted of two components: the emptying signal (the DC component), and a phasic 3 cycles per minutes (CPM) signal (the AC component). In some of the infants the phasic 3 CPM signal was also seen during the fasting state. For mature infants the median half emptying time (T50) was 6.9 min. For a second meal given within one hour after the first meal the half emptying time was 5.5 min (P < 0.01). In preterm infants the emptying times were not significantly different from mature infants. Day-to-day variation was low with a coefficient of variation of 17% in nine infants. A periodic change of the impedance signal, the phasic 3 CPM signal, was observed after a meal in 24 of the infants. The median frequency was 3.0 CPM in 20 mature and 2.9 CPM in four preterm infants. In nine infants a phasic 3 CPM signal was also observed during the fasting state, with a median frequency of 2.9 CPM. Measurement of gastric emptying pattern with epigastric impedance is a simple investigation for the evaluation of gastric emptying time and phasic activity in mature and preterm infants. However, the method is sensitive to spontaneous movements of the children, resulting in non-valid measurements in around one fourth of the infants.

Electric Impedance