PubMed · 8599299
A simple quality control protocol for haemoximetry using standard p50.
Abstract
We have designed a quality control tool for haemoximetry using the principle that standard p50, calculated from blood gas and haemoximetry measurements using the Siggaard-Andersen algorithm, remains unchanged when a specimen of venous blood is oxygenated progressively to 97% saturation of haemoglobin. If the haemoximeter is inaccurate, progressively larger deviations of standard p50 from the constant value occur above 80% saturation. Deviations too small to be detected by analysis of quality control materials comprised of non-haemoglobin dyes and yet sufficient to cause significant errors in derived extractivity parameters can be detected by this method. The suggested protocol is easy to perform, and is recommended as a quality control tool for blood gas analyser/haemoximeters used to determine arterial extractivity parameters.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
J Morgan, A Clague, J Rayfield, Z Endre. 1995. A simple quality control protocol for haemoximetry using standard p50.. https://pubmed.health/record/pubmed%3A8599299/a-simple-quality-control-protocol-for-haemoximetry-using-standard-p50
Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite PubMed Health by Indiciae: BibTeX · RIS.