Thyroid testing using the Cobas Core immunoassay system. A multicentre study.

Abstract

The random access immunoanalyzer Cobas Core and the Cobas Core Thyroid Assays were assessed as to their clinical usefulness in the analysis of different thyroid diseases. Four centres participated in this study measuring the following five thyroid tests per sample on the instrument: thyrotropin, free thyroxine, thyroxine, triiodothyronine and free triiodothyronine. The assessment was based on studies of precision and clinical samples. Within- and between-series precisions showed a mean CV over all assays of 4.3 and 6.1%, respectively. Comparison of the test results with clinical data demonstrated that the Cobas Core results are in accordance with diagnosed thyroid diseases. A good discrimination between normal and disease status and between untreated and treated status was found. Furthermore, 370 sera measured with Technicon Immuno-1 TSH and FT4 assays were compared to the respective Cobas Core EIAs). A good correlation between the assays was demonstrated.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

O Michotey, R Dörr, D Bourderont, J Gaya, F Rivera, R Benz, M Schmidt. 1995. Thyroid testing using the Cobas Core immunoassay system. A multicentre study.. https://pubmed.ncbi.nlm.nih.gov/8611671/

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Herniography and ultrasonography. A prospective study comparing the effectiveness of laparoscopic hernia repair with extraperitoneal balloon dissection.

BACKGROUND: This study was designed to assess differences between pre- and postoperative herniography and ultrasonography in inguinal hernia performed laparoscopically with balloon dissection and mesh without suture. METHODS: Pre- and postoperative herniographic and ultrasonographic findings were analyzed in ten consecutive patients. Postoperative ultrasonography was performed on the 3rd and 7th days and herniography was performed on the 7th day. RESULTS: Following the operation both the herniography and ultrasonography were almost normalized to a great extent in nine patients. Overall, minimal impaired continence was recorded by herniography in one patient. Also, we detected nonspecific soft-tissue thickening at the operation site in ultrasonographic examination in four patients. CONCLUSIONS: As for inguinal hernias, compared with other operative modalities of treatment, laparoscopic hernia repair with extraperitoneal balloon dissection and mesh without suture is a highly successful procedure and its minimal morbidity is well accepted by the patient.

Evaluation Studies as Topic

Splenic blood flow: evaluation with computed tomography.

RATIONALE AND OBJECTIVES: To study splenic perfusion with use of computed tomography (CT). METHODS: Twenty-six control patients without splenoportal disease, six with cirrhosis, and seven with other splenic disease were examined with electron-beam CT. Twenty-five milliliters of iohexol (300 mg of iodine per milliliter) was given intravenously at 10 mL/sec followed by a saline bolus. Multiple single-level axial sections were acquired 8-90 seconds after injection. Perfusion was calculated by dividing maximal splenic enhancement by the area under the circulation-corrected aortic time-enhancement curve. Subjective assessments of enhancement heterogeneity were made, and regional perfusion was calculated in 10 patients with heterogeneous enhancement. Total splenic volume and blood flow were computed in 21 patients. RESULTS: Mean perfusion (controls: 1.29 mL/min/mL, miscellaneous group: 1.07 mL/min/mL) was close to predictions. There was a trend toward lower perfusion in cirrhotic patients (0.87 mL/min/mL), but the difference was not statistically significant. Total splenic blood was increased in patients with cirrhosis (P < .01). Marked perfusion heterogeneity was observed in 41% of spleens, but by 2 minutes splenic enhancement was uniform. CONCLUSION: CT shows promise in the study of splenic blood flow.

Evaluation Studies as Topic