Friction and stem stiffness affect dynamic interface motion in total hip replacement.

Abstract

Large cyclic movements between the femoral stem and bone during the first weeks after total hip arthroplasty may hamper bone ingrowth and adversely affect the eventual success of the arthroplasty. Little is known, however, about the magnitude of the motions and its relationship to design and surgical factors. A two-dimensional finite element model of a cementless prosthesis inserted into the proximal femur was constructed to study the effects of two mechanical variables--the stiffness of the implant and the coefficient of friction between bone and implant--on the magnitude of the motions. We investigated the influences of these variables on the subsidence of the prosthesis, the magnitudes of the cyclic motions, and the level of the interface stresses. The presence of friction reduced cyclic motions by about 85% compared with a frictionless interface. Once friction was assumed, varying the coefficient of friction had little effect. The effect of friction on the interface stress state and gross subsidence of the prosthesis was not as great as on cyclic motion. Implant stiffness also affected the magnitudes and distributions of the cyclic motions along the interface. A flexible stem generated motions about three to four times larger proximally than those of a stiff stem, which generated larger motions distally. The influence of stem stiffness on interface stresses and prosthetic subsidence was less than on cyclic motion. The location of the peak shear stresses at the interface around a bonded prosthesis corresponded to the location where cyclic interface motion was maximal for an unbonded prosthesis. However, no direct relationship was found between the magnitudes of peak stresses and the amplitudes of cyclic motions.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J H Kuiper, R Huiskes. 1996. Friction and stem stiffness affect dynamic interface motion in total hip replacement.. https://doi.org/10.1002/jor.1100140108

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Dynamic mechanical analysis of polymeric systems of pharmaceutical and biomedical significance.

Dynamic mechanical analysis (DMA) is an analytical technique in which an oscillating stress is applied to a sample and the resultant strain measured as functions of both oscillatory frequency and temperature. From this, a comprehensive knowledge of the relationships between the various viscoelastic parameters, e.g. storage and loss moduli, mechanical damping parameter (tan delta), dynamic viscosity, and temperature may be obtained. An introduction to the theory of DMA and pharmaceutical and biomedical examples of the use of this technique are presented in this concise review. In particular, examples are described in which DMA has been employed to quantify the storage and loss moduli of polymers, polymer damping properties, glass transition temperature(s), rate and extent of curing of polymer systems, polymer-polymer compatibility and identification of sol-gel transitions. Furthermore, future applications of the technique for the optimisation of the formulation of pharmaceutical and biomedical systems are discussed.

Elasticity

Effect of magnesium stearate on bonding and porosity expansion of tablets produced from materials with different consolidation properties.

The negative effect of magnesium stearate on tablet strength is widely known. This strength reduction is always considered to be the result of reduction of interparticle bonding. It is also known that interparticle bonding affects relaxation of tablets. Relaxation increases with decreasing bonding. Microcrystalline cellulose is an example of a material with a high lubricant sensitivity, which effect is caused by its plastic deformation behavior during compression. This paper shows for microcrystalline cellulose that the porosity under pressure was equal for unlubricated tablets and for tablets containing 0.5% magnesium stearate. This points to equal densification properties. The lubricated tablets show, however, a much larger relaxation than the tablets without magnesium stearate. This difference can be ascribed to the reduction of interparticle bonding by the lubricant, because a strong interparticle bonding counteracts tablet relaxation. In contrast to microcrystalline cellulose, aggregated gamma-sorbitol (Karion Instant) has a low lubricant sensitivity. Both porosity under pressure and tablet relaxation were found to be equal for lubricated and unlubricated sorbitol tablets. This phenomenon is caused by the particle structure of gamma-sorbitol. During compression, a lubricant film will be destroyed by fragmentation of the sorbitol aggregates. For this reason, magnesium stearate will hardly affect the interparticle bonding between sorbitol particles and hence have only a small or no effect on tablet relaxation.

Elasticity

[Improved tumor stent for internal palliative urinary diversion].

The disadvantages of high flexible endoureteral stents (DJ) in case of tumorinduced extrinsic ureteral compression are due to an insufficient vertical stability of the used stents leading to stent-compression and consecutive hydro- or pyonephrosis. The new developed tumor-stent used in case of tumor-induced ureteral compression is available from 6 to 8 French in diameter and 24 to 32 cm in length. The corpus consists of a combination of high-stability plastics but is of sufficient elasticity in length. Both ends consist of extremely elastic J-parts guaranteeing an exact fixation. As against common DJ's with the same outside-diameter the new stent has a comparable interior diameter and compared to used "old" tumor stents promises a higher interior flow in case of extrinsic diseases. The application can be undertaken in well-known technique, needs no special instrumentation and no learning-curve. To date 52 stents at our urologic departments were placed without any problems, the latest remaining for 15 months. Tumor-induced compression or a higher rate of encrustation could not be seen. All patients tolerated these stents well. In our opinion the new stabilized endoureteral stent can be seen as a better solution instead of percutaneous nephrostomy or frequent stent changing in patients with tumor induced extrinsic ureteral compression.

Elasticity