Positive peak identification in liquid chromatography using absorbance ratioing with a variable-wavelength spectrophotometric detector.

Abstract

It is frequently necessary to confirm the identification of component peaks isolated in high-performance liquid chromatograms. The UV spectrum of the pure compound is usually a very powerful signature for most compounds. For liquid chromatographic instruments that permit stop-flow operation, a simple approximation is obtained with a variable wavelength UV detector by determining the absorbance ratios at several specific wavelengths to characterize the compound. If the wavelengths are well chosen, these absorbance ratios provide a very specific and a very reliable technique for peak identification. The absorbance ratios are independent of the concentration of the compounds in the detector flow cell. Identity or purity of individual peaks can be confirmed when absorbance ratios are taken from the peaks of individually injected pure standards. Flow can be stopped several times on the leading and trailing edges of broad peaks, and ratios calculated. If the ratios are identical, the peak can be considered to be pure; if ratios differ, the presence of unresolved components can be suspected. Stop-flow ratioing can be performed during gradient elution: standards are injected under rapid elution conditions and the resulting ratios are compared with those of the peaks from the gradient run. Ratios are found to be identical for similar components, regardless of the differences in retention time, thus providing rapid identification of the compounds in the gradient chromatogram. Factors that influence the overall precision of the method are: ability to stop and re-start eluent flow without loss of chromagraphic efficiency, ability to obtain numerical absorbance data from the detector, the wavelength repeatability of the detector, and the requirement that the detector cell design allow slow interruptions without baseline upset.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R Yost, J Stoveken, W Maclean. 1977-04-01. Positive peak identification in liquid chromatography using absorbance ratioing with a variable-wavelength spectrophotometric detector.. https://doi.org/10.1016/s0021-9673(00)82571-6

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

HPLC separation and NMR structural elucidation of sn-1,2-, 2,3-, and 1,3-diacylglycerols from olive oil as naphthylethylurethane derivatives.

In this study, sn-1,2-, sn-2,3-, and sn-1,3-diacylglycerols were isolated from olive oil, and their urethane derivatives (urethanes) were prepared. Normal-phase high-performance liquid chromatography (NP-HPLC) separation of the urethane isomers was performed and the separate classes were studied by nuclear magnetic resonance (NMR). The use of 1H NMR and homo- and heteronuclear 2D techniques provided a great amount of information in a very short time, particularly when a high-field NMR instrument (700 MHz) was used. Particularly diagnostic for this kind of compound was the glyceridic moiety that presents typical chemical shifts both for carbon and hydrogen. These studies show the usefulness of NMR spectroscopy to recognize clearly the sn-1,3- and, moreover, sn-1,2- with respect to sn-2,3-diacylglycerols, although very minor differences occur between them.

Chromatography, High Pressure Liquid↗

Solid-phase extraction combined with high-performance liquid chromatography-atmospheric pressure chemical ionization-mass spectrometry analysis of pesticides in water: method performance and application in a reconnaissance survey of residues in drinking water in Greater Cairo, Egypt.

Monitoring of water resources for pesticide residues is often needed to ensure that pesticide use does not adversely impact the quality of public water supplies or the environment. In many rural areas and throughout much of the developing world, monitoring is often constrained by lack of testing facilities; thus, collection of samples and shipment to centralized laboratories for analysis is required. The portability, ease of use, and potential to enhance analyte stability make solid-phase extraction (SPE) an attractive technique for handling water samples prior to their shipment. We describe performance of an SPE method targeting a structurally diverse mixture of 25 current-use pesticides and two common degradates in samples of raw and filtered drinking water collected in Greater Cairo, Egypt. SPE was completed in a field laboratory in Egypt, and cartridges were shipped to the United States for elution and high-performance liquid chromatography-atmospheric pressure chemical ionization-mass spectrometry analysis. Quantitative and reproducible recovery of 23 of 27 compounds (average = 96%; percent relative standard deviation = 21%) from matrix spikes (1 microg L-1 per component) prepared in the field and from deionized water fortified similarly in the analytical laboratory was obtained. Concurrent analysis of unspiked samples identified four parent compounds and one degradate in drinking water samples. No significant differences were observed between raw and filtered samples. Residue levels in all cases were below drinking water and "harm to aquatic-life" thresholds, indicating that human and ecological risks of pesticide contamination were relatively small; however, the study was limited in scale and scope. Further monitoring is needed to define spatial and temporal variation in residue concentrations. The study has demonstrated the feasibility of performing studies of this type using SPE to extract and preserve samples in the field. The approach should be broadly applicable in many settings.

Chromatography, High Pressure Liquid↗

Retention behavior of oligomeric proanthocyanidins in hydrophilic interaction chromatography.

A novel method was developed for the separation of proanthocyanidins (PAs; oligomeric flavan-3-ols) by hydrophilic interaction chromatography (HILIC) using an amide-silica column eluting with an aqueous acetonitrile mobile phase. The best separation was achieved with a linear gradient elution of acetonitrile-water at ratios of 9:1 to 5:5 (v/v) for 60 min at a flow rate of 1.0 ml/min. Under these HPLC conditions, a mixture of natural oligomeric PAs (from apple) was separated according to degree of polymerization (DP) up to decamers. The DP of each separated oligomer was confirmed by LC/electrospray ionization MS. In further HILIC separation studies of 15 different flavan-3-ol and oligomeric PA (up to pentamer) standards with an isocratic elution of acetonitrile-water (84:16), a high correlation was observed between the logarithm of retention factors (log k) and the number of hydroxyl groups in their structures. The coefficient of this correlation (r2=0.9501) was larger than the coefficient (r2=0.7949) obtained from the correlation between log k and log P(o/w) values. These data reveal that two effects, i.e. hydrogen bonding between the carbamoyl terminal on the column and the hydroxyl group of solute oligomer and hydrophilicity based on the high-order structure of oligomeric PAs, corporately contribute to the separation, but the hydrogen bonding effect is predominant in our HILIC separation mode.

Chromatography, High Pressure Liquid↗