PubMed · 8830626
Paraprotein interference causing pseudohyperphosphataemia: evaluation of an improved methodology.
Abstract
The measurement of serum inorganic phosphorus by direct acid/molybdate assay is known to be subject to interference by some serum paraproteins. Serum samples from 45 paraproteinaemic patients were used for the measurement of inorganic phosphorus using a manufacturer's kit reagent and a detergent-modified reagent from the same manufacturer. In addition, the same samples were analysed with both reagents following precipitation of serum protein with trichloroacetic acid. Using the original reagent pseudohyperphosphataemia was found in 27% (12/45) of the samples. The modified reagent was free from interference with no significant difference between results from serum and protein-free supernatants.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
D J Savory, C J Pearce. 1995. Paraprotein interference causing pseudohyperphosphataemia: evaluation of an improved methodology.. https://doi.org/10.1177/000456329503200511
Cite the original work for its findings. Save a collection to share your selection of sources.