Computed tomography artifacts associated with craniofacial fixation devices: an experimental study.

Abstract

This study compares the artifacts caused by eight different craniofacial fixation devices in computed tomography (CT) images. Using a Teflon CT phantom model, part I of this study involved the quantitative evaluation of the X-ray absorption properties of each fixation device. Part II utilized a human cadaveric model to determine the degree to which the artifact interfered with the visualization of anatomic structures. In part I, each fixation device was secured to the surface of the phantom and then scanned. All artifacts were compared on the basis of standard deviation in CT number. The severity of the artifact was related to the physical size of the fixation device and its composition. Vitallium devices generated a greater degree of CT artifacts than titanium devices of comparable size. In part II, fixation devices were secured to the orbital rims of human cadaveric heads and then scanned. Visualization of specified anatomic structures was graded independently. The results revealed that titanium fixation devices did not cause significant bone or soft-tissue image degradation, whereas all vitallium fixation devices, except micro mesh and micro (1.0 mm) straight plates, generated an artifact that resulted in some image degradation. The extent of image degradation was related to the fixation device size. Only the thickest vitallium fixation device, mini fragmentation (2.0 mm), resulted in bony image degradation. The degree of soft-tissue image degradation decreased as the size of vitallium fixation devices decreased such that micro fragmentation (0.8 mm) and pan fixation (1.3 mm) devices interfered with soft-tissue visualization only in the immediate vicinity of the plate. The results of this study confirm the previous work of Sullivan and colleagues and Fiala and associates. The data indicate that when postoperative imaging is an important clinical consideration: (1) the fewest number of internal fixation devices should be used to achieve rigid bony fixation, (2) the proximity of fixation devices to the regions of interest should be considered at the time of fixation, (3) titanium implants produce less artifacts than vitallium implants of comparable size, and (4) vitallium micro mesh and micro (1.0 mm) straight fixation devices do not produce artifacts resulting in significant image degradation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D J Anastakis, O M Antonyshyn, P W Cooper, M J Yaffe, K Bush, G E Mawdsley. 1996. Computed tomography artifacts associated with craniofacial fixation devices: an experimental study.. https://doi.org/10.1097/00000637-199610000-00002

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Striping artifact removal in VisiumHD data through nuclear counts modeling.

MOTIVATION: 10x Genomics VisiumHD enables spatial transcriptomics at 2 µm × 2 µm resolution but exhibits slide-specific, non-periodic striping artifacts due to lane-width variability. These multiplicative row/column effects distort bin total counts and can bias downstream analyses. The state-of-the-art destriping approach is the normalization procedure used as a preprocessing step in bin2cell; it applies sequential high-quantile row- then column-wise normalization, which is asymmetric and can introduce edge effects/macro-stripes and distortions of large-scale total-count structure. RESULTS: We propose a statistical destriping approach that leverages nuclei segmentation from the co-registered H&E image. Assuming transcript abundance is constant within each nucleus, we model bin counts with a negative binomial distribution whose mean is a product of a nucleus-specific concentration and row- and column-specific stripe-factors reflecting lane-width variation. We fit all parameters in a generalized linear modeling framework with cross-validated regularization on stripe-factors and iterative dispersion estimation, and use the fitted parameters to correct the observed counts into a destriped image. On synthetic data with known ground truth, our method improves stripe-factor estimation accuracy and reduces error in corrected counts relative to bin2cell and bin2cell-derived baselines. Across four public VisiumHD slides, it consistently lowers striping intensity while substantially better preserving biological signal present in the large-scale global count structure and avoiding the artifacts introduced by other methods. AVAILABILITY AND IMPLEMENTATION: All source code and links to publicly available data used for this study are available at https://github.com/paolamalsot/destriping-GLM.

Artifacts↗

Impairment of echocardiographic acoustic window caused by breast implants.

Cosmetic breast implants are increasing in popularity. The presence of a breast implant overlying the anterior mediastinal space as a cause of impairment of the echocardiographic acoustic window has not been described previously. Here, we report three cases with significant impairment of echocardiographic acoustic window caused by breast implants. Clinicians should be aware of this interference and women should be informed of this dilemma before considering this cosmetic surgery.

Artifacts↗