A method to reduce systematic spatial shift associated with magnetic resonance imaging.
Abstract
To reduce the chemical shifts during magnetic resonance (MR) imaging, the authors replaced the petroleum gel in the Brown-Roberts-Well (BRW) MR localizer with chromium chloride. Computed tomography and MR scans were obtained of a phantom skull containing objects with known spatial coordinates. A 2-to 3-mm systematic spatial shift in the frequency-encoded direction was observed with petroleum gel, but not with CrCl3. Results were verified by reconstructing the three-dimensional spatial location of each object using X-Knife computer software. The authors conclude that spatial localization is more accurate with a CrCl3-filled than a petroleum-filled BRW-MR localizer.