PubMed · 900068
A rapid screening method for blood glucose estimation.
Abstract
A rapid method for estimating glucose concentrations in serum and in fluoride-, iodoacetate-, heparin-, and EDTA-treated plasma and whole blood is described. The procedure requires about three minutes to perform and utilize a tungstic acid precipitant solution and urine glucose dipsticks. Test results correlate with those of a reference quantitative glucose method at levels from 25 to 500 mg/dl (1.38 to 27.5 mmol/l). Hemolysis, lipemia and bilirubin levels as high as 20 mg/dl (342 mmol/l) do not interfere with the procedure. The simplicity and adaptability of the method make it useful in emergency situations.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
W L Gyure. 1977. A rapid screening method for blood glucose estimation.. https://doi.org/10.1093/ajcp%2F68.3.336
Cite the original work for its findings. Save a collection to share your selection of sources.