MRI using piecewise-linear spiral trajectory.

Abstract

A new generation of high power gradient systems which allow much faster MR imaging as well as shorter echo times has recently become available. Some of these high-speed gradient systems impose limits on the percentage of time during which the gradient can change in amplitude (slewing duty cycle). While this limitation may be immaterial to many 2DFT and echo planar imaging methods, a traditional circular spiral trajectory is difficult to use on these systems because its gradient waveforms change during the entire course of the trajectory so that the slewing duty cycle during the readout period is 100%. We describe a piecewise-linear spiral trajectory which is composed of linear segments and rounded corners. This trajectory reduces the slewing duty cycle while maintaining the desirable imaging properties of circular spirals including interleaving by simple gradient rotation. For one representative example, the slewing duty cycle is reduced to 46%. A conventional gridding method was used for image reconstruction, but a new numerical algorithm to calculate the density compensation factor was required. Use of piecewise-linear spiral trajectories reduces the impact imposed by limited gradient slewing duty cycle.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J R Liao, J M Pauly, N J Pelc. 1997. MRI using piecewise-linear spiral trajectory.. https://doi.org/10.1002/mrm.1910380213

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A method for determining B1 field inhomogeneity. Are the biases assumed in heteronuclear relaxation experiments usually underestimated?

We describe a method allowing the determination of the effective B1 field amplitude distribution in a high-resolution NMR spectrometer. This method which can be adapted to almost any sequence, essentially consists of a mutation followed by a purging B0 gradient pulse. Experimental results obtained with this approach are described in homonuclear and heteronuclear cases. The experimental distributions are used to estimate the biases induced by B1 inhomogeneity, as well as the loss of RF power on heteronuclear transverse self-relaxation rate determination. In this type of measurement, the experimental biases induced on the intensities can be as large as 5% for long mixing times.

Image Processing, Computer-Assisted

J-multiplied HSQC (MJ-HSQC): a new method for measuring 3J(HNHalpha) couplings in 15N-labeled proteins.

A new method for the measurement of homonuclear 3J(HNHalpha) coupling constants in 15N-labeled small proteins is described. The method is based on a modified sensitivity enhanced HSQC experiment, where the 3J(HNHalpha) couplings are multiplied in the f1-dimension. The J-multiplication of homonuclear 3J(HNHalpha) couplings is based on simultaneous incrementation of 15N chemical shift and homonuclear coupling evolution periods. The time increment for the homonuclear coupling evolution period is chosen to be a suitable multiple (2N x t1) of the corresponding increment for 15N-shift evolution. This results in the splitting of the HSQC correlation in the f1-dimension by 2N x 3J(HNHalpha). Because the pulse sequence has good sensitivity and water suppression properties, it is particularly useful for natural abundance samples.

Image Processing, Computer-Assisted

Three dimensional image restoration in fluorescence lifetime imaging microscopy.

A microscope set-up and numerical methods are described which enable the measurement and reconstruction of three-dimensional nanosecond fluorescence lifetime images in every voxel. The frequency domain fluorescence lifetime imaging microscope (FLIM) utilizes phase detection of high-frequency modulated light by homodyne mixing on a microchannel plate image intensifier. The output signal at the image intensifier's phosphor screen is integrated on a charge coupled device camera. A scanning stage is employed to obtain a series of phase-dependent intensity images at equally separated depths in a specimen. The Fourier transform of phase-dependent data gives three-dimensional (3D) images of the Fourier coefficients. These images are deblurred using an Iterative Constrained Tikhonov-Miller (ICTM) algorithm in conjunction with a measured point spread function. The 3D reconstruction of fluorescence lifetimes are calculated from the deblurred images of the Fourier coefficients. An improved spatial and temporal resolution of fluorescence lifetimes was obtained using this approach to the reconstruction of simulated 3D FLIM data. The technique was applied to restore 3D FLIM data of a live cell specimen expressing two green fluorescent protein fusion constructs having distinct fluorescence lifetimes which localized to separate cellular compartments.

Image Processing, Computer-Assisted