PubMed · 9404582

Instrument performance in bone density testing at five Australian centres.

Abstract

AIMS: To assess the in vitro precision and accuracy of bone mineral densitometry (BMD) within and between locations at five Australian centres. METHODS: Using a multicentre reliability study the accuracy and short- and long-term precision of dual-energy X-ray absorptiometry (DXA) in vitro was compared on five instruments. Measures were performed using pencil beam mode on four Hologic QDR-2000 densitometers and one Hologic QDR-1000/W (Hologic Inc, Waltham, MA). RESULTS: Short-term precision of bone mineral density measurement was less than 0.5% for spine phantoms (n = 10 for each centre, mean intrasite coefficient of variation [CV] 0.39 +/- 0.09% [SD]) and for hip phantoms (n = 10 for each centre, mean intrasite coefficient of variation [CV] 0.34 +/- 0.10% [SD]). Between-centre measurement (n = 10 for each phantom) of a single spine phantom and a single hip phantom (specified mineral contents-58.5 g and 38.6 g, respectively) revealed ranges of bone mineral content of 57.7-58.1 g (all-point CV = 0.52%) and 37.1-37.8 g (all-point CV = 0.70%), respectively. When results from pairs of machines were compared there were statistically different mean BMD results for the majority of the ten possible pairings for both spine and hip measurements. Each study centre measured in vitro stability of phantom BMD measurements over a one year period (n = 45-283, median 157 for spine; and n = 0-262, median 38, for hip); CVs ranged from 0.38 to 0.53% for the spine measurements and from 0.38 to 0.54% for the hip measurements. The mean all-point accuracy of the spine phantom measurements was 99.1% and the hip phantom measurements was 96.7%. CONCLUSIONS: Across a number of instruments DXA demonstrates in vitro all-point precision of 0.5% for the spine phantom and 0.7% for the hip phantom. The instrument demonstrates accuracy of greater than 99% at the spine and 96% at the hip. This finding has clinical, research and quality control implications.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

K M Khan, S L Henzell, C Broderick, R L Prince, A Saul, J Lomman, J D Wark. 1997. Instrument performance in bone density testing at five Australian centres.. https://doi.org/10.1111/j.1445-5994.1997.tb00959.x

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

A repeated measures concordance correlation coefficient.

The concordance correlation coefficient is commonly used to assess agreement between two raters or two methods of measuring a response when the data are measured on a continuous scale. However, the situation may arise in which repeated measurements are taken for each rater or method, e.g. longitudinal studies in clinical trials or bioassay data with subsamples. This paper proposes a coefficient for measuring agreement between two raters or two methods of measuring a response in the presence of repeated measurements. We illustrate the methodology with examples comparing (1) 1-hr versus 2-hr blood draws for measuring cortisol in an asthma clinical trial and (2) two measurements of percentage body fat, from skinfold calipers and dual energy X-ray absorptiometry.

Absorptiometry, Photon↗

Crystallization and preliminary X-ray analysis of a domain in the Runx2 transcription factor that interacts with the 1alpha,25 dihydroxy vitamin D3 receptor.

The Runx2 transcription factor is a key regulator of osteoblast differentiation. In response to 1alpha,25 dihydroxy vitamin D3, Runx2 may interact with the 1alpha,25 dihydroxy vitamin D3 receptor (VDR) in the promoter of target genes, producing a synergic activation of their transcription. Previous studies have suggested that the motifs responsible for the VDR-Runx2 interaction are contained within the 230-361 domain of Runx2. In this work, we confirmed by GST-pull down that Runx2(I(209-361)) is sufficient to interact with the VDR. To obtain structural information, GST-Runx2(I(209-361)) protein was overexpressed in Escherichia coli, purified and crystallized using the hanging-drop vapor-diffusion method and polyethyleneglycol as a precipitant. The crystals were found to diffract to a maximum resolution of 2.7 A and a complete data set to a 3.3 A resolution was collected and analyzed. The crystals belong to the tetragonal system, with a space group P4 and unit-cell parameters of a = b = 90.8, and c = 57.2 A. The presence of a monomer of the recombinant GST-Runx2(I(209-361)) in the asymmetric unit gives a V(M) of 2.7 A(3) Da(-1) and a solvent content of 54.8%.

Absorptiometry, Photon↗

Quantifying the treatment effect explained by markers in the presence of measurement error.

Surrogate markers or intermediate markers are important in identifying subjects with high risk of a serious disease or for monitoring disease progression of a subject on treatment. Quantifying the proportion of treatment effect (PTE) explained by markers has been studied extensively. Due to reasons such as biological variation, limited machine precision, etc. markers are generally measured with error. The estimated PTE ignoring the measurement error could be biased, which may lead to incorrect conclusions. In this article, we adjust for the measurement error using regression calibration to construct a less biased estimator of excess relative odds, a quantity to measure the treatment effect explained by markers. The method is applied to data from a clinical study in osteoporosis.

Absorptiometry, Photon↗