PubMed · 9404830
Errors of computer electrocardiography.
Abstract
A review of 6938 computer-generated electrocardiogram interpretations was undertaken by a single cardiologist. The computer-assisted program correctly interpreted 51.6% of the tracings. One or more corrections were necessary in 48.4%. The greatest number of incorrect interpretations involved repolarization (T wave) abnormalities, followed by ST segment changes. It is mandatory that computer-generated ECG interpretations be reviewed by an experienced electrocardiographer. Reliance on nonreviewed ECG interpretations may result in incorrect diagnoses that could lead to inappropriate management decisions.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
B M RuDusky. 1997. Errors of computer electrocardiography.. https://doi.org/10.1177/000331979704801204. https://pubmed.health/record/pubmed%3A9404830/errors-of-computer-electrocardiography