Composite hip prosthesis design. I. Analysis.

Abstract

An investigation was performed to study the mechanical performance of fiber-reinforced composite hip prostheses in a femur. The main objective of the study was to evaluate the effect of fiber orientation of a composite femoral implant on the response of the surrounding femoral bone. A three-dimensional finite element analysis was developed for analyzing a composite implant in the femur. A three-dimensional composite element was proposed to take into account ply drop-off due to a change of cross-section of the composite implant. The element could accommodate multidirectional layers and tapered composites. The material properties of the composite were treated as anisotropic and inhomogeneous while the properties of femoral bone were treated as anisotropic and homogenous. All the materials were assumed to behave linear-elastically. The thermoplastic graphite/PEEK material system was selected for the study. In this presentation, as the first part of the study, the development of the finite element analysis will be described. Numerical calculations were generated and compared with existing data and numerical results available from studies related to metal hip prostheses in the literature. Experiments on the composite hip implants were also conducted for further verification of the analysis and the computer simulations. In Part II, using the finite element code, an extensive study was performed to evaluate the stress/strain distributions, micromotions, and strain-energy density of the surrounding femoral bone, which have been related to initial fixation and long-term stability of the prosthesis in a femoral bone. Numerous fiber orientations were studied, and results of the calculations were compared with those generated by the prosthesis made of cobalt alloy and titanium alloy.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

H Yildiz, S K Ha, F K Chang. 1998. Composite hip prosthesis design. I. Analysis.. https://doi.org/10.1002/(sici)1097-4636(199801)39%3A1%3C92%3A%3Aaid-jbm12%3E3.0.co%3B2-q

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

The use of megavoltage cone-beam CT to complement CT for target definition in pelvic radiotherapy in the presence of hip replacement.

In Europe and the USA combined, over half a million people had a hip joint replaced in 2005, contributing to the increasing number of radiotherapy patients with metallic hip prostheses. The treatment plan for external beam radiation therapy is based on the delineation of the anatomy in the planning CT scan. When implanted objects of high atomic number (Z) material are present, however, severe image artefacts are generated in conventional CT, strongly hindering the ability to delineate some organs. This is particularly the case for the planning of prostate patients with hip prostheses. This short communication presents the use of a new imaging modality, megavoltage cone-beam CT, to complement the regular CT for target definition of prostate cancer treatment of patients with hip replacements.

Hip Prosthesis↗

Stem damage during implantation of modular hip prostheses.

Modular hip prostheses are commonly used today. Dedicated surgical auxiliary instrumentation is indispensable. A correct coupling between the prosthesis and the instrument is necessary to avoid damage. Based on a specific commercial design (AnCAFit, Wright Cremascoli Ortho, Toulon, France), the aim of this work was to assess if stem holders of modular prostheses can induce any stem damage, and if so, to identify the features that provide minimal damage. Two different stem holder connector designs were investigated. The surface of the stem cavity was inspected to identify the damage caused by the use of the two connectors after a simulated surgical handling. Both connectors induced some damage, with significantly more damage being caused by a supposedly improved design. Scars due to contact with the stem holder were observed on the stem cavity surface. This may be sufficient to trigger increased fretting damage during service life, causing the production of fretting debris.

Hip Prosthesis↗