Anchoring coil embolization in a high-flow arterial model: a pilot study.
Abstract
PURPOSE: To devise and test an occluding coil anchoring system to improve the safety of coil embolization. MATERIALS AND METHODS: The anchoring system was attached to Gianturco embolization coils and investigated in 15 pigs. In the short-term studies, one 0.035-inch anchored coil (15-18 cm in length and 7-10 mm in diameter) was placed in the infrarenal portion of the abdominal aorta in each of 12 pigs with use of an 8-F catheter from the carotid approach. Aortography was performed before and up to 4 hours after coil placement. In the long-term studies, 0.028-inch anchored coils (8 cm in length and 5 mm in diameter) were placed in the left femoral and the right carotid arteries in each of three pigs with use of a 6-F catheter positioned from the right femoral approach. One week later, the animals were evaluated angiographically for coil migration and vascular occlusion. RESULTS: Radiographically, the coils created a compact conglomerate on placement in all but one of the animals. No coil migration was noted during follow-up. Necropsy confirmed compact arrangement of the coils within the vessels and revealed effective anchoring of the device in all cases. CONCLUSIONS: The anchoring coil has proved effective in making coil embolization safer, especially in a high-flow arterial model.