Anaesthesia circuits, humidity output, and mucociliary structure and function.

Abstract

We compared the effects of humidity delivered by the circle system at low fresh gas flows (FGF) with a conventional two-limb and coaxial circuit on the structure and function of the tracheobronchial epithelium in dogs. Animals were anaesthetized and mechanically ventilated using an anaesthesia ventilator to maintain normocarbia. Group I (control) animals received a FGF equal to the required minute ventilation mimicking an open circuit technique. Group II and III animals had FGF set at 20% of the required minute ventilation. Group II used a two-limb circuit and Group III used a coaxial circuit. Relative humidity and temperature of inspired gases were measured at baseline and hourly afterwards. In the first experiment, biopsies of the tracheobronchial tree were obtained bronchoscopically at baseline and then hourly for six hours. Microscopic examination of these samples allowed calculation of mean ciliary length. In the second experiment, tracheal mucus flow velocity (TMFV) was measured at baseline and hourly afterward, using a cinebroncho-fibrescopic method. Delivered absolute humidity was greatest with low FGF and the coaxial circuit, followed by low FGF and a conventional circuit, and high FGF (15 +/- 1.4 vs 9 +/- 0.8 vs 5 +/- 0.4 mg H2O, P < 0.01) after two hours. Mean cilia length (micron) and TMFV (mm/min) fell during the first hour in all three groups. At hour two TMFV returned to baseline in Group III and was significantly greater than Groups I and II (0.8 +/- 0.4 vs 8.6 +/- 1.1 vs 15.4 +/- 2.1, P < 0.001). Mean ciliary length demonstrated a similar pattern with reductions from baseline in all three groups for the first two hours. Groups II and III had an increase in cilia length beginning at hour three and were both significantly greater than Group I at hours 3 through 6 (1.3 +/- 0.5 vs 3.2 +/- 1.1 vs 4.2 +/- 0.8, P < 0.001). Alterations in tracheobronchial structure and function result from exposure to dry gases and are amplified by the duration of exposure. Our findings suggest a minimum of 12 to 15 mg H2O/l is necessary to prevent these alterations. In this study, the combination of low FGF and a coaxial anaesthesia circuit reached this minimum threshold more quickly than a conventional two-limb circuit.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

R D Branson, R S Campbell, K Davis, D T Porembka. 1998. Anaesthesia circuits, humidity output, and mucociliary structure and function.. https://doi.org/10.1177/0310057x9802600208

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related citations

Minimal access thoracic surgery in the pediatric population.

Thoracoscopy was initially described for use in children to obtain pulmonary biopsy samples in the immunocompromised patient. With refinements in technique, development of better instrumentation, and advances in pediatric anesthesia, there are now many diagnostic and therapeutic indications for the use of thoracoscopy in children. One of the most common indications includes pleural debridement for empyema. Many centers consider this the optimal approach for biopsy of mediastinal lesions and excision of bronchogenic or duplication cysts. The technique is useful for pleural disorders, such as spontaneous pneumothorax and chylothorax. Thoracoscopy has been used to achieve exposure for spinal diskectomy in children with thoracic scoliosis, and newer techniques are being developed in performing anatomic lobectomies, repair of esophageal atesias, and closure of diaphragmatic hernias. The role of the robot in pediatric thoracoscopy is still in the early stages of definition.

Anesthesia, Closed-Circuit↗

Closed system anaesthesia--historical aspects and recent developments.

Closed circuit anaesthesia was described decades ago but did not achieve wide popularity among anaesthesiologists mainly because reliable control of inspiratory gas concentrations was not possible. Recent innovations including fast gas analysers, electronically controlled dosage systems and algorithms for feedback control have made possible the development of sophisticated closed circuit ventilators designed for routine clinical practice. The main advantages comprise economic use of medical gases and volatile anaesthetics, reduction of anaesthetic gas loss into the atmosphere, improved airway acclimatization as well as estimations of oxygen consumption. This article reviews historical aspects, recent developments as well as advantages and limitations of closed system anaesthesia.

Anesthesia, Closed-Circuit↗